Nakajima Erika, Sato Hitoshi
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki Prefecture, Japan.
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki Prefecture, Japan.
Z Med Phys. 2025 May;35(2):169-176. doi: 10.1016/j.zemedi.2023.05.004. Epub 2023 Jun 24.
Radiochromic film (RCF) is a detector that can obtain a two-dimensional dose distribution with high resolution; it is widely used in medical and industrial fields. Several types of RCFs exist based on their application. The type of RCF mainly used for mammography dose assessment has been discontinued; however, a new type of RCF (LD-V1) has been distributed as a successor. Since the medical use of LD-V1 has rarely been studied, we investigated the response characteristics of LD-V1 in mammography.
Measurements were performed using Mo/Mo and Rh/Ag on a Senographe Pristina mammography device (GE, Fairfield, CT, USA). The reference air kerma was measured using a parallel-plate ionization chamber (PPIC) (C-MA, Applied Engineering Inc, Tokyo, Japan). Pieces of LD-V1 film model were irradiated at the same position where the reference air kerma in air was measured by the PPIC. Irradiation was performed using the time scale method based on the load on the equipment. Two methods of irradiation were considered: placing the detector in air and on the phantom. The LD-V1 was scanned five times at 72 dpi in RGB (48 bit) mode using a flatbed scanner (ES-G11000, Seiko Epson Corp, Nagano, Japan) 24 h following irradiation. The response ratio of the reference air kerma and the air kerma obtained from the LD-V1 were compared and examined for each beam quality and air kerma range.
When the beam quality was altered, the response ratio varied from 0.8 to 1.2 with respect to the measurement value of the PPIC; however, some outliers were observed. Response ratios were highly variable in the low-dose range; however, as the air kerma increased, the ratios approached 1. Thus, LD-V1 does not need calibration for each beam quality used in mammography. LD-V1 enables air kerma evaluation by creating air kerma response curves under certain X-ray conditions used in mammography.
We suggest that the dose range be limited to 12 mGy or more to keep the response variation with beam qualities below ±20%. If further measurement is required for reducing the response variation, the dose range should be shifted to a higher dose range.
放射变色薄膜(RCF)是一种能够高分辨率获取二维剂量分布的探测器,在医学和工业领域广泛应用。基于其应用存在几种类型的RCF。主要用于乳腺摄影剂量评估的RCF类型已停产;然而,一种新型的RCF(LD-V1)作为其替代品已开始销售。由于对LD-V1的医学用途研究较少,我们研究了LD-V1在乳腺摄影中的响应特性。
在美国康涅狄格州费尔菲尔德市通用电气公司的Senographe Pristina乳腺摄影设备上,使用钼/钼和铑/银进行测量。使用平行板电离室(PPIC)(日本东京应用工程公司的C-MA)测量参考空气比释动能。将LD-V1薄膜模型片放置在PPIC测量空气中参考空气比释动能的相同位置进行照射。照射采用基于设备负载的时间刻度法。考虑了两种照射方法:将探测器置于空气中和置于模体上。照射后24小时,使用平板扫描仪(日本长野精工爱普生公司的ES-G11000)以72 dpi的分辨率在RGB(48位)模式下对LD-V1进行五次扫描。比较并检查了每种射线质和空气比释动能范围内参考空气比释动能与从LD-V1获得的空气比释动能的响应比。
当射线质改变时,相对于PPIC的测量值,响应比在0.8至1.2之间变化,但观察到一些异常值。在低剂量范围内响应比变化很大;然而,随着空气比释动能增加,这些比值接近1。因此,LD-V1不需要针对乳腺摄影中使用的每种射线质进行校准。LD-V1能够通过在乳腺摄影中使用的特定X射线条件下创建空气比释动能响应曲线来评估空气比释动能。
我们建议将剂量范围限制在12 mGy或更高,以使射线质引起的响应变化保持在±20%以下。如果需要进一步测量以减少响应变化,剂量范围应转移到更高剂量范围。