Thorax Center, Erasmus MC, Rotterdam, The Netherlands.
J Biomed Opt. 2012 Jul;17(7):076018. doi: 10.1117/1.JBO.17.7.076018.
Quantitative analysis of optical coherence tomography data can be strongly hampered by speckle. Here, we introduce a new method to reduce speckle, which leverages from Fourier-domain configurations and operates on individual axial scans. By subdividing the digitized spectrum into a number of distinct narrower windows, each with a different center frequency, several independent speckle patterns result. These can be averaged to yield a lower-resolution image with strongly reduced speckle. The full resolution image remains available for human interpretation; the low resolution version can be used for parametric imaging or quantitative analysis. We demonstrate this technique using intravascular optical frequency domain imaging data acquired in vivo.
光学相干断层扫描(OCT)数据的定量分析可能会受到散斑的严重阻碍。在这里,我们介绍了一种新的方法来减少散斑,该方法利用了傅里叶域配置,并对单个轴向扫描进行操作。通过将数字化光谱分为多个不同的较窄窗口,每个窗口具有不同的中心频率,会产生多个独立的散斑图案。可以对这些图案进行平均,以生成具有强烈减少的散斑的较低分辨率图像。完整的分辨率图像仍然可供人解释;低分辨率版本可用于参数成像或定量分析。我们使用体内血管内光学频域成像(IVOFI)数据演示了该技术。