Department of Human Oncology, University of Wisconsin, Madison, WI, USA.
Radiol Oncol. 2011 Sep;45(3):220-6. doi: 10.2478/v10019-011-0024-5. Epub 2011 Jul 20.
We have analyzed the stability of CT to density curve of kilovoltage cone-beam computerized tomography (kV CBCT) imaging modality over the period of six months. We also, investigated the viability of using image value to density table (IVDT) generated at different time, for adaptive radiotherapy treatment planning. The consequences of target volume change and the efficacy of kV CBCT for adaptive planning issues is investigated. MATERIALS AND METHODS.: Standard electron density phantom was used to establish CT to electron density calibrations curve. The CT to density curve for the CBCT images were observed for the period of six months. The kV CBCT scans used for adaptive planning was acquired with an on-board imager system mounted on a "Trilogy" linear accelerator. kV CBCT images were acquired for daily setup registration. The effect of variations in CT to density curve was studied on two clinical cases: prostate and lung.
The soft tissue contouring is superior in kV CBCT scans in comparison to mega voltage CT (MVCT) scans. The CT to density curve for the CBCT images was found steady over six months. Due to difficulty in attaining the reproducibility in daily setup for the prostate treatment, there is a day-to-day difference in dose to the rectum and bladder.
There is no need for generating a new CT to density curve for the adaptive planning on the kV CBCT images. Also, it is viable to perform the adaptive planning to check the dose to target and organ at risk (OAR) without performing a new kV CT scan, which will reduce the dose to the patient.
我们分析了千伏锥形束 CT(kV CBCT)成像模式的 CT 到密度曲线在六个月期间的稳定性。我们还研究了在不同时间生成的图像值到密度表(IVDT)用于自适应放射治疗计划的可行性。研究了靶区变化的后果和 kV CBCT 对自适应计划问题的有效性。
标准电子密度体模用于建立 CT 到电子密度校准曲线。观察了 CBCT 图像的 CT 到密度曲线六个月。用于自适应计划的 kV CBCT 扫描是使用安装在“Trilogy”直线加速器上的机载成像系统获取的。kV CBCT 图像用于每日设置注册。研究了 CT 到密度曲线变化对两个临床病例(前列腺和肺)的影响。
kV CBCT 扫描中的软组织轮廓优于兆伏 CT(MVCT)扫描。CBCT 图像的 CT 到密度曲线在六个月内保持稳定。由于难以在前列腺治疗中实现每日设置的可重复性,直肠和膀胱的剂量存在每日差异。
kV CBCT 图像的自适应计划不需要生成新的 CT 到密度曲线。此外,在不进行新的 kV CT 扫描的情况下,执行自适应计划以检查靶区和危及器官(OAR)的剂量是可行的,这将降低患者的剂量。