Influence of oxygenator type on the prevalence and extent of microembolic retinal ischemia during cardiopulmonary bypass. Assessment by digital image analysis.

作者信息

Blauth C I, Smith P L, Arnold J V, Jagoe J R, Wootton R, Taylor K M

机构信息

Department of Surgery, RPMS, Hammersmith Hospital, London, England.

出版信息

J Thorac Cardiovasc Surg. 1990 Jan;99(1):61-9.

PMID:2294366
Abstract

We have previously reported the occurrence of microembolic ischemia in the retina during cardiopulmonary bypass, as revealed by fluorescein angiography. This method has been extended by digital image analysis to include quantification of the extent of retinal ischemia and has been applied to a prospective comparative study of 64 patients undergoing elective coronary operations with either a bubble or a membrane oxygenator. Patients with diabetes or clinically evident cerebrovascular disease were excluded. Bypass procedures were standardized in all cases with pulsatile flow and a 40 microns arterial line filter (Pall EC Plus). Thirty patients had bypass with a bubble oxygenator (Harvey H1700) and 34 patients had bypass with a flat sheet membrane oxygenator (COBE CML). In each case retinal fluorescein angiograms were obtained preoperatively and 5 minutes before the end of bypass and were processed with a digital image analyzer (Context Vision GOP-302). Microembolic perfusion defects were identified by digital subtraction of preoperative and end-bypass angiograms and their total area was computed. Results. In the bubble oxygenator group retinal perfusion defects indicative of microembolism occurred in all 30 (100%) patients. In contrast, over half the patients in the membrane oxygenator group had normal retinal perfusion, and the prevalence of perfusion defects (44%; 70% confidence limits 34% to 54%) was significantly less than in the bubble group (p less than 0.001). In addition, those patients in the membrane group had significantly fewer lesions (median 0; 70% confidence limits 0 to 1) than patients in the bubble group (median 2; 70% confidence limits 2 to 2; p less than 0.001) and also had significantly smaller total areas of retinal ischemia (median 0 mm2; 70% confidence limits 0 to 0.16 mm2) than the bubble group (median 0.22 mm2; 70% confidence limits 0.21 to 0.27 mm2; p less than 0.001). There was no relationship between the extent of retinal ischemia and bypass time, arterial blood gas concentrations, volume of cardiotomy suction or donor blood returned to the pump, or recent medication with aspirin. Inferences. Digital image analysis of retinal fluorescein angiograms may provide a method of quantifying microembolic ischemia in the central nervous system during cardiopulmonary bypass. Flat sheet membrane oxygenation appears to provide significantly better protection against microembolic ischemia than bubble oxygenation.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验