Glüer C C, Steiger P, Selvidge R, Elliesen-Kliefoth K, Hayashi C, Genant H K
Department of Radiology, University of California, San Francisco 94143.
Radiology. 1990 Jan;174(1):223-8. doi: 10.1148/radiology.174.1.2294552.
Dual-energy bone densitometry can be performed with two types of scanners. The traditional dual-photon absorptiometry (DPA) machines use an isotope source, whereas the newly introduced dual-energy radiography (DER) devices use an incorporated x-ray tube. The authors evaluated the performance of DPA and DER devices at their institution. The short-term precision error in vivo was 1.2% for femoral neck measurements with DER. Long-term precision error in vitro was reduced from 1.30% (DPA) to 0.44% (DER). The scanning time for both spine and hip measurements was reduced from 20-40 minutes to 6-7 minutes. Intraosseous fat sensitivity remained the same, at a level of 12 mg/cm2 apparent decrease of bone mineral density (BMD) per 10% fat by volume change, and for both devices there was no shift in BMD when phantom thickness was increased by 1.5 inches. The correlation of DPA and DER was high: r = .98 for the spine and r = .95 for the femoral neck. Correlation of DPA versus quantitative computed tomography (CT) (r = .83) and DER versus quantitative CT (r = .85) was good. The advent of DER represents a significant advance for the field of bone densitometry.
双能骨密度测定可通过两种类型的扫描仪进行。传统的双光子吸收法(DPA)机器使用同位素源,而新推出的双能X线摄影(DER)设备则使用内置的X射线管。作者评估了他们机构中DPA和DER设备的性能。DER测量股骨颈的体内短期精度误差为1.2%。体外长期精度误差从1.30%(DPA)降至0.44%(DER)。脊柱和髋部测量的扫描时间从20 - 40分钟减少到6 - 7分钟。骨内脂肪敏感性保持不变,每10%体积的脂肪变化导致骨矿物质密度(BMD)表观下降12mg/cm²,并且当体模厚度增加1.5英寸时,两种设备的BMD均无变化。DPA和DER的相关性很高:脊柱的r = 0.98,股骨颈的r = 0.95。DPA与定量计算机断层扫描(CT)的相关性(r = 0.83)以及DER与定量CT的相关性(r = 0.85)都很好。DER的出现代表了骨密度测定领域的一项重大进展。