Evaluation of the Sysmex R-1000. An automated reticulocyte analyzer.

作者信息

Tichelli A, Gratwohl A, Driessen A, Mathys S, Pfefferkorn E, Regenass A, Schumacher P, Stebler C, Wernli M, Nissen C

机构信息

Department of Laboratory Medicine, University Hospital, Basel, Switzerland.

出版信息

Am J Clin Pathol. 1990 Jan;93(1):70-8. doi: 10.1093/ajcp/93.1.70.

Abstract

The new fully automated reticulocyte analyzer, Sysmex R-1000 (TOA Medical Electronics, Kobe, Japan), was evaluated for its routine use in the Hematological Laboratory at the University Hospital Basel, Switzerland. The operating characteristics, such as within-run precision, linearity, and carryover, fulfilled the manufacturer's specifications and are excellent. Correlation with the standard method, manual reticulocyte counting, is linear for normal and high values. For low reticulocyte counts the regression points show a deviation from their linearity. An absolute zero value is not obtained by the R-1000. The R-1000 measures total RNA content of each cell and expresses the value as low fluorescence ratio (LFR), medium fluorescence ratio (MFR), and high fluorescence ratio (HFR). The analysis of this ratio resolves the problem of zero reticulocytes: A fraction of less than 0.002 (0.2%) with an LFR of 100% represents aplasia; a shift of the intensity of fluorescence to HFR heralds regeneration. Results of samples stored at room temperature remain stable and within the range of the within-run precision for up to 12 hours, when stored at 5 degrees C for more than 48 hours. The authors conclude that the R-1000 is easy to operate, fulfills the criteria for accuracy and precision, and is highly suitable for daily routine use in a large central hematologic laboratory.

摘要

新型全自动网织红细胞分析仪Sysmex R - 1000(日本神户东亚医用电子公司)在瑞士巴塞尔大学医院血液学实验室接受了常规使用评估。其运行特性,如批内精密度、线性度和携带污染率,均符合制造商的规格要求,表现出色。与标准方法(手工网织红细胞计数)相比,在正常和高值情况下呈线性相关。对于低网织红细胞计数,回归点显示出与线性关系的偏差。R - 1000无法获得绝对零值。R - 1000测量每个细胞的总RNA含量,并将该值表示为低荧光率(LFR)、中荧光率(MFR)和高荧光率(HFR)。对该比率的分析解决了零网织红细胞的问题:LFR为100%且比例小于0.002(0.2%)表示再生障碍;荧光强度向HFR的转变预示着再生。室温下保存的样本结果在长达12小时内保持稳定,且在批内精密度范围内,若在5摄氏度下保存超过48小时也是如此。作者得出结论,R - 1000易于操作,符合准确性和精密度标准,非常适合大型中心血液学实验室的日常常规使用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验