Department of Ophthalmology, Hospital Universitario Miguel Servet, Zaragoza, Spain.
Am J Ophthalmol. 2013 Jan;155(1):171-176.e1. doi: 10.1016/j.ajo.2012.06.012. Epub 2012 Sep 8.
To determine the interobserver and intraobserver reproducibility of a Fourier-domain optical coherence tomography device (Cirrus HD OCT; Carl Zeiss Meditec, Dublin, California, USA) in normal pediatric eyes.
Prospective cross-sectional study.
One hundred healthy children were recruited prospectively and consecutively. Only 1 randomly chosen eye per subject was included in the study. The eye underwent 3 scans centered on the optic disc and another 3 scans centered on the macula that were acquired by a single operator. A fourth examination was performed by a second operator. Interobserver and intraobserver reproducibility were described by intraclass correlation coefficients (ICCs) and coefficients of variation (COVs).
The mean age was 9.15 years (range, 6.22 to 11.31 years; standard deviation, 1.05 years). Mean retinal nerve fiber layer thickness was 99.53 μm (standard deviation, 10.10 μm), and mean macular thickness was 282.91 μm (standard deviation, 11.83 μm). All the parameters evaluated were highly reproducible. Intraobserver COVs of the retinal nerve fiber layer measurements ranged from 2.24% to 5.52%, and the COV of macular thickness was 0.97%. The intraclass correlation coefficient was greater than 0.8 for all the parameters. The interobserver COV ranged from 2.23% to 5.18%, and the COV of macular thickness was 0.82%. In all the evaluated parameters, the intraclass correlation coefficient was more than 0.75. Repeatability was slightly better in children older than 10 years than in children younger than 9 years.
Retinal nerve fiber layer and macular measurements obtained by Fourier-domain optical coherence tomography showed good repeatability for healthy eyes in the pediatric population. Cirrus HD OCT examinations of the retina are reliable in children.
确定傅里叶域光学相干断层扫描设备(Cirrus HD OCT;美国加利福尼亚州都柏林卡尔蔡司医疗技术公司)在正常儿科眼中的观察者间和观察者内可重复性。
前瞻性横断面研究。
前瞻性连续招募 100 名健康儿童。每个受试者仅随机选择一只眼纳入研究。该眼接受了 3 次以视盘为中心的扫描和 3 次以黄斑为中心的扫描,均由一名操作人员完成。由第二名操作人员进行第四次检查。观察者间和观察者内的可重复性通过组内相关系数(ICC)和变异系数(COV)来描述。
平均年龄为 9.15 岁(范围,6.22 至 11.31 岁;标准差,1.05 岁)。平均视网膜神经纤维层厚度为 99.53μm(标准差,10.10μm),平均黄斑厚度为 282.91μm(标准差,11.83μm)。所有评估的参数均具有高度可重复性。视网膜神经纤维层测量的观察者内 COV 范围为 2.24%至 5.52%,黄斑厚度的 COV 为 0.97%。所有参数的 ICC 均大于 0.8。观察者间 COV 范围为 2.23%至 5.18%,黄斑厚度的 COV 为 0.82%。在所有评估的参数中,ICC 均大于 0.75。10 岁以上儿童的可重复性略优于 9 岁以下儿童。
在儿科人群中,傅里叶域光学相干断层扫描获得的视网膜神经纤维层和黄斑测量值具有良好的重复性。Cirrus HD OCT 对儿童视网膜的检查是可靠的。