Department of Radiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China.
Clin Radiol. 2013 Mar;68(3):272-8. doi: 10.1016/j.crad.2012.08.007. Epub 2012 Sep 13.
To investigate image quality and potential for radiation dose reduction using sinogram-affirmed iterative reconstruction (SAFIRE) at computed tomography (CT) coronary angiography (CTCA) compared with filtered back-projection (FBP) reconstruction.
A water phantom and 49 consecutive patients were scanned using a retrospectively electrocardiography (ECG)-gated CTCA protocol on a dual-source CT system. Image reconstructions were performed with both conventional FBP and SAFIRE. The SAFIRE series were reconstructed image data from only one tube, simulating a 50% radiation dose reduction. Two blinded observers independently assessed the image quality of each coronary segment using a four-point scale and measured image noise (the standard deviation of Hounsfield values, SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Radiation dose estimates were calculated.
In the water phantom, image noise decreased at the same ratio as the tube current increased for both reconstruction algorithms. Despite an estimated radiation dose reduction from 7.9 ± 2.8 to 4 ± 1.4 mSv, there was no significant difference in the SD and SNR within the aortic root and left ventricular chamber between the two reconstruction methods. There was also no significant difference in the image quality between the FBP and SAFIRE series.
Compared with traditional FBP, there is potential for substantial radiation dose reduction at CTCA with use of SAFIRE, while maintaining similar diagnostic image quality.
与滤波反投影(filtered back-projection,FBP)重建相比,通过体层摄影证实迭代重建(sinogram-affirmed iterative reconstruction,SAFIRE)评估 CT 冠状动脉成像(CTCA)的图像质量和降低辐射剂量的潜力。
使用双源 CT 系统对水模体和 49 例连续患者进行回顾性心电门控 CTCA 扫描。采用常规 FBP 和 SAFIRE 进行图像重建。SAFIRE 系列是通过仅使用一个管模拟 50%辐射剂量减少重建的图像数据。两名盲法观察者使用四点量表独立评估每个冠状动脉节段的图像质量,并测量图像噪声(Hounsfield 值的标准差,SD)、信噪比(signal-to-noise ratio,SNR)和对比噪声比(contrast-to-noise ratio,CNR)。计算辐射剂量估计值。
在水模体中,两种重建算法的图像噪声均随管电流的增加以相同的比例降低。尽管估计的辐射剂量从 7.9±2.8 降至 4±1.4 mSv,但两种重建方法在主动脉根部和左心室腔的 SD 和 SNR 之间没有显著差异。FBP 和 SAFIRE 系列之间的图像质量也没有显著差异。
与传统 FBP 相比,使用 SAFIRE 进行 CTCA 可显著降低辐射剂量,同时保持相似的诊断图像质量。