Yang Wen Jie, Yan Fu Hua, Liu Bo, Pang Li Fang, Hou Liang, Zhang Huan, Pan Zi Lai, Chen Ke Min
Department of Radiology, Rui Jin Hospital, Medical School, Shanghai Jiao Tong University, Shanghai, PR China.
J Comput Assist Tomogr. 2013 Mar-Apr;37(2):301-5. doi: 10.1097/RCT.0b013e31827b8c66.
To evaluate the performance of sinogram-affirmed iterative (SAFIRE) reconstruction on image quality of low-dose lung computed tomographic (CT) screening compared with filtered back projection (FBP).
Three hundred four patients for annual low-dose lung CT screening were examined by a dual-source CT system at 120 kilovolt (peak) with reference tube current of 40 mA·s. Six image serials were reconstructed, including one data set of FBP and 5 data sets of SAFIRE with different reconstruction strengths from 1 to 5. Image noise was recorded; and subjective scores of image noise, images artifacts, and the overall image quality were also assessed by 2 radiologists.
The mean ± SD weight for all patients was 66.3 ± 12.8 kg, and the body mass index was 23.4 ± 3.2. The mean ± SD dose-length product was 95.2 ± 30.6 mGy cm, and the mean ± SD effective dose was 1.6 ± 0.5 mSv. The observation agreements for image noise grade, artifact grade, and the overall image quality were 0.785, 0.595 and 0.512, respectively. Among the overall 6 data sets, both the measured mean objective image noise and the subjective image noise of FBP was the highest, and the image noise decreased with the increasing of SAFIRE reconstruction strength. The data sets of S3 obtained the best image quality scores.
Sinogram-affirmed iterative reconstruction can significantly improve image quality of low-dose lung CT screening compared with FBP, and SAFIRE with reconstruction strength 3 was a pertinent choice for low-dose lung CT.
与滤波反投影(FBP)相比,评估正弦图确认迭代(SAFIRE)重建对低剂量肺部计算机断层扫描(CT)筛查图像质量的影响。
304例接受年度低剂量肺部CT筛查的患者,采用双源CT系统,在120千伏(峰值)、参考管电流40毫安·秒的条件下进行检查。重建了6组图像序列,包括1组FBP数据集和5组不同重建强度(1至5)的SAFIRE数据集。记录图像噪声;2名放射科医生还对图像噪声、图像伪影和整体图像质量进行了主观评分。
所有患者的平均体重±标准差为66.3±12.8千克,体重指数为23.4±3.2。平均剂量长度乘积±标准差为95.2±30.6毫希沃特·厘米,平均有效剂量±标准差为1.6±0.5毫希沃特。图像噪声等级、伪影等级和整体图像质量的观察一致性分别为0.785、0.595和0.512。在全部6组数据集中,FBP的测量平均客观图像噪声和主观图像噪声均最高,且图像噪声随SAFIRE重建强度的增加而降低。S3数据集获得了最佳图像质量评分。
与FBP相比,正弦图确认迭代重建可显著提高低剂量肺部CT筛查的图像质量,重建强度为3的SAFIRE是低剂量肺部CT的合适选择。