Suppr超能文献

从一个常见的假定健康人群中确定 19 个肌钙蛋白 I 和 T 检测的第 99 个百分位值。

Determination of 19 cardiac troponin I and T assay 99th percentile values from a common presumably healthy population.

机构信息

Hennepin County Medical Center Department of Laboratory Medicine and Pathology, Minneapolis, MN 55415, USA.

出版信息

Clin Chem. 2012 Nov;58(11):1574-81. doi: 10.1373/clinchem.2012.192716. Epub 2012 Sep 14.

Abstract

BACKGROUND

Between-assay comparability of 99th percentiles for cardiac troponin concentrations has not been assessed systematically in a single population for a large number of assays.

METHODS

We determined 99th percentiles for 19 cardiac troponin assays in heparin plasma samples from a population of 272 and 252 presumably healthy males and females, respectively. The assays evaluated included 1 cardiac troponin T (cTnT) assay from Roche and 18 cTnI assays from Abbott, Alere, Beckman, bioMerieux, Instrumentation Laboratory, Ortho-Clinical Diagnostics, Singulex, Siemens, and Roche. Five of these assays were categorized as high-sensitivity, 9 as sensitive-contemporary, and 5 as point-of-care (POC) assays.

RESULTS

For high-sensitivity cTnI (hs-cTnI) assays 99th percentiles varied from 23 to 58 ng/L. At least 80% of individuals had measurable hs-cTnI, whereas only 25% had measurable high-sensitivity cTnT. All high-sensitivity cardic troponin assays had 99th percentiles that were 1.2-2.4-fold higher in males than females. For the 9 sensitive-contemporary cTnI assays, 99th percentiles varied from 12 to 392 ng/L, and only the Beckman assay gave measurable concentrations in a substantial portion of the population (35% vs ≤6% for the others). Seven of these 9 assays had 1.3-5.0-fold higher 99th percentiles for males than females. For 5 cTnI POC assays, 99th percentiles varied from <10 to 40 ng/L. The Instrumentation Laboratory assay gave measurable results in 27.8% of study participants vs ≤6% for the others. Correlations were generally poor among assays.

CONCLUSIONS

Among cardiac troponin assays 99th percentile concentrations appear to differ. High-sensitivity assays provide measurable cardiac troponin results in a substantially greater fraction of presumably healthy individuals.

摘要

背景

在单一人群中,尚未针对大量检测方法对心脏肌钙蛋白浓度的第 99 百分位值进行系统的检测间可比性评估。

方法

我们在分别由 272 名和 252 名疑似健康男性和女性的肝素血浆样本中确定了 19 种心脏肌钙蛋白检测方法的第 99 百分位值。评估的检测方法包括罗氏的 1 种心脏肌钙蛋白 T(cTnT)检测和雅培、艾利尔、贝克曼、伯乐、仪器实验室、奥森多临床诊断、辛加莱克斯、西门子和罗氏的 18 种心脏肌钙蛋白 I(cTnI)检测。其中 5 种被归类为高敏,9 种为敏感-当前,5 种为即时检测(POC)检测。

结果

对于高敏心脏肌钙蛋白 I(hs-cTnI)检测,第 99 百分位值范围为 23 至 58ng/L。至少 80%的个体具有可测量的 hs-cTnI,而只有 25%的个体具有可测量的高敏心脏肌钙蛋白 T。所有高敏心脏肌钙蛋白检测在男性中的第 99 百分位值均比女性高 1.2-2.4 倍。对于 9 种敏感-当前心脏肌钙蛋白 I 检测,第 99 百分位值范围为 12 至 392ng/L,只有贝克曼检测在很大一部分人群中(35%比其他检测的 6%)具有可测量的浓度。其中 7 种检测在男性中的第 99 百分位值比女性高 1.3-5.0 倍。对于 5 种心脏肌钙蛋白 POC 检测,第 99 百分位值范围为<10 至 40ng/L。仪器实验室检测在 27.8%的研究参与者中得到了可测量的结果,而其他检测的结果<6%。检测之间的相关性通常较差。

结论

在心脏肌钙蛋白检测中,第 99 百分位值浓度似乎存在差异。高敏检测方法在更多数量的疑似健康个体中提供可测量的心脏肌钙蛋白结果。

文献AI研究员