Department of Cardiology, University of Gießen, Klinikstr. 33, 35392 Gießen, Germany.
Eur Radiol. 2013 Mar;23(3):597-606. doi: 10.1007/s00330-012-2656-2. Epub 2012 Sep 16.
We evaluated the potential of prospectively ECG-triggered high-pitch spiral acquisition with low tube voltage and current in combination with iterative reconstruction to achieve coronary CT angiography with sufficient image quality at an effective dose below 0.1 mSv.
Contrast-enhanced coronary dual source CT angiography (2 × 128 × 0.6 mm, 80 kV, 50 mAs) in prospectively ECG-triggered high-pitch spiral acquisition mode was performed in 21 consecutive individuals (body weight <100 kg, heart rate ≤60/min). Images were reconstructed with raw data-based filtered back projection (FBP) and iterative reconstruction (IR). Image quality was assessed on a 4-point scale (1 = no artefacts, 4 = unevaluable).
Mean effective dose was 0.06 ± 0.01 mSv. Image noise was significantly reduced in IR (128.9 ± 46.6 vs. 158.2 ± 44.7 HU). The mean image quality score was lower for IR (1.9 ± 1.1 vs. 2.2 ± 1.0, P < 0.0001). Of 292 coronary segments, 55 in FBP and 40 in IR (P = 0.12) were graded "unevaluable". In patients with a body weight ≤75 kg, both in FBP and in IR, the rates of fully evaluable segments were significantly higher in comparison to patients >75 kg.
Coronary CT angiography with an estimated effective dose <0.1 mSv may provide sufficient image quality in selected patients through the combination of high-pitch spiral acquisition and raw data-based iterative reconstruction.
我们评估了前瞻性心电图触发高螺旋采集结合低管电压和电流以及迭代重建在有效剂量低于 0.1mSv 的情况下实现冠状动脉 CT 血管造影的潜力,以获得足够的图像质量。
在 21 例连续个体(体重<100kg,心率≤60/min)中进行前瞻性心电图触发高螺旋采集模式下的对比增强冠状动脉双源 CT 血管造影(2×128×0.6mm,80kV,50mAs)。使用原始数据滤波反投影(FBP)和迭代重建(IR)重建图像。使用 4 分制评估图像质量(1=无伪影,4=无法评估)。
平均有效剂量为 0.06±0.01mSv。IR 时图像噪声显著降低(128.9±46.6 与 158.2±44.7HU)。IR 的平均图像质量评分较低(1.9±1.1 与 2.2±1.0,P<0.0001)。292 个冠状动脉节段中,FBP 有 55 个,IR 有 40 个(P=0.12)评为“无法评估”。在体重≤75kg 的患者中,FBP 和 IR 中完全可评估节段的比例均明显高于体重>75kg 的患者。
通过高螺旋采集和原始数据迭代重建的结合,在选择的患者中,估计有效剂量<0.1mSv 的冠状动脉 CT 血管造影可以提供足够的图像质量。