LDR 近距离治疗¹⁹²Ir 线材的剂量学研究:蒙特卡罗和 TPS 计算。

Dosimetric investigation of LDR brachytherapy ¹⁹²Ir wires by Monte Carlo and TPS calculations.

机构信息

Department of Physics, Faculty of Arts and Sciences, Harran University, Osmanbey Campus, 63300 Sanliurfa, Turkey.

出版信息

Jpn J Radiol. 2013 Jan;31(1):24-30. doi: 10.1007/s11604-012-0140-1. Epub 2012 Oct 10.

Abstract

PURPOSE

The aim of this study was to investigate the dose rate distribution around (192)Ir wires used as radioactive sources in low-dose-rate brachytherapy applications.

MATERIALS AND METHODS

Monte Carlo modeling of a 0.3-mm diameter source and its surrounding water medium was performed for five different wire lengths (1-5 cm) using the MCNP software package.

RESULTS

The computed dose rates per unit of air kerma at distances from 0.1 up to 10 cm away from the source were first verified with literature data sets. Then, the simulation results were compared with the calculations from the XiO CMS commercial treatment planning system.

CONCLUSION

The study results were found to be in concordance with the treatment planning system calculations except for the shorter wires at close distances.