Department of Medical Radiation Physics, Lund University, Lund, Sweden.
J Magn Reson Imaging. 2012 Dec;36(6):1450-9. doi: 10.1002/jmri.23778. Epub 2012 Oct 12.
To evaluate accuracy and noise properties of a novel time-resolved, three-dimensional, three-directional phase contrast sequence with variable velocity encoding (denoted 4D-vPC) on a 3 Tesla MR system, and to investigate potential benefits and limitations of variable velocity encoding with respect to depicting blood flow patterns.
A 4D PC-MRI sequence was modified to allow variable velocity encoding (VENC) over the cardiac cycle in all three velocity directions independently. 4D-PC sequences with constant and variable VENC were compared in a rotating phantom with respect to measured velocities and noise levels. Additionally, comparison of flow patterns in the ascending aorta was performed in six healthy volunteers.
Phantom measurements showed a linear relationship between velocity noise and velocity encoding. 4D-vPC MRI presented lower noise levels than 4D-PC both in phantom and in volunteer measurements, in agreement with theory. Volunteer comparisons revealed more consistent and detailed flow patterns in early diastole for the variable VENC sequences.
Variable velocity encoding offers reduced noise levels compared with sequences with constant velocity encoding by optimizing the velocity-to-noise ratio (VNR) to the hemodynamic properties of the imaged area. Increased VNR ratios could be beneficial for blood flow visualizations of pathology in the cardiac cycle.
在 3T 磁共振系统上评估一种新型的时间分辨、三维、三方向相位对比序列(记为 4D-vPC)的准确性和噪声特性,并研究可变速度编码在描绘血流模式方面的潜在优势和局限性。
对 4D PC-MRI 序列进行了修改,使其能够在整个心动周期内独立地在三个速度方向上进行可变速度编码(VENC)。在旋转体模中,对具有恒定和可变 VENC 的 4D-PC 序列进行了比较,比较了测量速度和噪声水平。此外,还对六名健康志愿者的升主动脉血流模式进行了比较。
体模测量结果表明,速度噪声与速度编码之间存在线性关系。4D-vPC MRI 无论是在体模还是志愿者测量中,其噪声水平均低于 4D-PC,这与理论相符。志愿者比较结果显示,在可变 VENC 序列中,舒张早期的血流模式更一致、更详细。
与具有恒定速度编码的序列相比,可变速度编码通过将速度-噪声比(VNR)优化到成像区域的血液动力学特性,可以降低噪声水平。增加 VNR 比值可能有助于在心脏周期中对病理的血流可视化。