经食管下括约肌电刺激治疗 GERD 效果良好:开放性前瞻性试验的最终结果。

Electrical stimulation therapy of the lower esophageal sphincter is successful in treating GERD: final results of open-label prospective trial.

机构信息

Department of Surgery, Centro Clínico de Obesidad (CCO), Obesidad Y Diabetes, Estoril N° 120 Of. 814, Las Condes, Santiago, Chile.

出版信息

Surg Endosc. 2013 Apr;27(4):1083-92. doi: 10.1007/s00464-012-2561-4. Epub 2012 Oct 17.

Abstract

BACKGROUND

Electrical stimulation of the lower esophageal sphincter (LES) improves LES pressure without interfering with LES relaxation. The aim of this open-label pilot trial was to evaluate the safety and efficacy of long-term LES stimulation using a permanently implanted LES stimulator in patients with gastroesophageal reflux disease (GERD).

METHODS

GERD patients who were at least partially responsive to proton pump inhibitors (PPI) with abnormal esophageal pH, hiatal hernia ≤ 3 cm, and esophagitis ≤ LA grade C were included. Bipolar stitch electrodes were placed in the LES and an IPG was placed in a subcutaneous pocket. Electrical stimulation was delivered at 20 Hz, 215 μs, 3-8 mA in 30 min sessions. The number and timing of sessions was tailored to each patient's GERD profile. Patients were evaluated using GERD-HRQL, daily symptom and medication diaries, SF-12, esophageal pH, and high-resolution manometry.

RESULTS

24 patients (mean age = 53 years, SD = 12 years; 14 men) were implanted; 23 completed their 6-month evaluation. Median GERD-HRQL scores at 6 months was 2.0 (IQR = 0-5.5) and was significantly better than both baseline on-PPI [9.0 (range = 6.0-10.0); p < 0.001] and off-PPI [23 (21-25); p < 0.001] GERD-HRQL. Median% 24-h esophageal pH < 4.0 at baseline was 10.1 and improved to 5.1 at 6 months (p < 0.001). At their 6-month follow-up, 91 % (21/23) of the patients were off PPI and had significantly better median GERD-HRQL on LES stimulation compared to their on-PPI GERD-HRQL at baseline (9.0 vs. 2.0; p < 0.001). There were no unanticipated implantation- or stimulation-related adverse events or untoward sensation due to stimulation. There were no reports of treatment-related dysphagia, and manometric swallow was also unaffected.

CONCLUSIONS

Electrical stimulation of the LES is safe and effective for treating GERD. There is a significant and sustained improvement in GERD symptoms, esophageal pH, and reduction in PPI usage without any side effects with the therapy. Furthermore, the therapy can be optimized to address an individual patient's disease.

摘要

背景

刺激食管下括约肌(LES)可改善 LES 压力,而不会干扰 LES 松弛。本开放性试验研究的目的是评估使用永久性植入式 LES 刺激器对胃食管反流病(GERD)患者进行长期 LES 刺激的安全性和疗效。

方法

本研究纳入了质子泵抑制剂(PPI)至少部分反应、食管 pH 值异常、食管裂孔疝≤3cm 和食管炎≤LA C 级的 GERD 患者。双极缝合电极放置在 LES 中,IPG 放置在皮下袋中。以 20Hz、215μs、3-8mA 的强度进行 30 分钟的刺激。每个患者的刺激时间和强度都根据 GERD 情况进行调整。通过 GERD-HRQL、每日症状和药物日记、SF-12、食管 pH 值和高分辨率测压法对患者进行评估。

结果

24 例患者(平均年龄=53 岁,标准差=12 岁;14 例男性)接受了植入治疗;23 例完成了 6 个月的评估。6 个月时 GERD-HRQL 中位数为 2.0(IQR=0-5.5),明显优于 PPI 治疗时的基线值[9.0(范围=6.0-10.0);p<0.001]和 PPI 停药时的基线值[23(21-25);p<0.001]。基线时 24 小时食管 pH 值<4.0 的中位数为 10.1%,6 个月时改善至 5.1%(p<0.001)。在 6 个月的随访中,91%(21/23)的患者停用了 PPI,与基线时 PPI 治疗时的 GERD-HRQL 相比,LES 刺激时的 GERD-HRQL 明显更好(9.0 比 2.0;p<0.001)。无植入或刺激相关不良事件,也无因刺激引起的不良反应。没有因治疗引起的吞咽困难报告,测压吞咽也不受影响。

结论

LES 刺激治疗 GERD 安全有效。GERD 症状、食管 pH 值显著改善,PPI 使用减少,且无任何副作用,治疗效果可持续。此外,该疗法可针对患者的个体疾病进行优化。