Body composition in subjects with anorexia nervosa: bioelectrical impedance analysis and dual-energy X-ray absorptiometry.

机构信息

Dipartimento di Sanità Pubblica, Università degli Studi di Firenze, Viale GB Morgagni 48, 50134 Firenze, Italy.

出版信息

Eat Weight Disord. 2012 Dec;17(4):e298-303. doi: 10.3275/8722. Epub 2012 Nov 14.

Abstract

OBJECTIVE

To evaluate the applicability and validity of bioelectrical impedance analysis (BIA) compared with dual-energy X-ray absorptiometry (DXA) in a population of girls with restrictive anorexia nervosa (AN).

METHODS

A total of 30 girls (11-19 years old) with AN were enrolled. DXA and BIA (BIA software and the Deurenberg equations) were used to estimate the body composition. The correlation between the methods was assessed by Pearson's correlation coefficient and the Bland-Altman method.

RESULTS

The mean FFM estimates were 33.2 kg (BIA software), 32.8 kg (BIA, Deurenberg equation) and 33.1 kg (DXA). The mean FM values were 5.6 kg (BIA software), 6.2 kg (BIA, Deurenberg equation) and 6.4 kg (DXA). There was a high correlation between the FFM values estimated with the two methods (BIA software vs DXA r=0.917, p<0.001; Deurenberg equation vs DXA r=0.931, p<0.001). For the FFM, the limits of agreement were equal to ±3.34 kg for the BIA software and ±2.96 kg for the Deurenberg equation. For the FM, the limits of agreement were equal to ±4.60 kg for the BIA software and ±3.82 kg for the Deurenberg equation.

CONCLUSION

The results show a good correlation between DXA and BIA. BIA seems to be a valid alternative for epidemiological and clinical evaluations.

摘要

目的

评估生物电阻抗分析(BIA)与双能 X 射线吸收法(DXA)在限制型厌食症(AN)女孩人群中的适用性和有效性。

方法

共纳入 30 名 AN 女孩(年龄 11-19 岁)。使用 DXA 和 BIA(BIA 软件和 Deurenberg 方程)来估计身体成分。通过 Pearson 相关系数和 Bland-Altman 法评估两种方法之间的相关性。

结果

FFM 的平均估计值分别为 BIA 软件 33.2kg、BIA,Deurenberg 方程 32.8kg 和 DXA 33.1kg。FM 的平均估计值分别为 BIA 软件 5.6kg、BIA,Deurenberg 方程 6.2kg 和 DXA 6.4kg。两种方法(BIA 软件与 DXA r=0.917,p<0.001;Deurenberg 方程与 DXA r=0.931,p<0.001)之间的 FFM 值相关性很高。对于 FFM,BIA 软件的一致性界限为±3.34kg,Deurenberg 方程的一致性界限为±2.96kg。对于 FM,BIA 软件的一致性界限为±4.60kg,Deurenberg 方程的一致性界限为±3.82kg。

结论

结果表明 DXA 与 BIA 之间具有良好的相关性。BIA 似乎是一种有效的替代方法,可用于流行病学和临床评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验