Pakravan Mohammad, Pakbin Mojgan, Aghazadehamiri Mohammad, Yazdani Shahin, Yaseri Mehdi
Ophthalmic Research Center and Department of Ophthalmology, Labbafinejad Medical Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Eur J Ophthalmol. 2013 May-Jun;23(3):289-95. doi: 10.5301/ejo.5000167. Epub 2012 Nov 23.
To determine agreement between peripapillary retinal nerve fiber layer (RNFL) thickness measurements by 2 spectral domain optical coherence tomography (SD-OCT) instruments.
. This was an observational, cross-sectional study. Both eyes of 89 subjects, 44 healthy subjects (mean age, 47±15 years) and 45 patients with glaucoma (mean age, 55±14 years), were imaged using Cirrus OCT (Carl Zeiss Meditec, Inc. Dublin, CA, USA) and Topcon OCT (3D-OCT 1000, Topcon Corporation, Tokyo, Japan). One eye was randomly selected in each subject and peripapillary RNFL thickness measurements were compared. Bland-Altman plots were used to assess the level of agreement.
. Average peripapillary RNFL thickness measurements were significantly larger with Topcon OCT compared to Cirrus OCT in normal (101±12 vs 91±8 µm, p<0.001) and glaucomatous eyes (89±18 vs 76±16 µm, p<0.001). The limits of agreement for average peripapillary RNFL thickness measurements were poor, (-11 to 32 µm) in normal and (-4 to 30 µm) in glaucomatous eyes.
. Peripapillary RNFL thickness values obtained by Topcon OCT are significantly higher than those measured by Cirrus OCT in both normal and glaucomatous eyes. Clinicians should be aware of this discrepancy especially when monitoring patients imaged by different OCT instruments; measured values cannot be used interchangeably.
确定两种光谱域光学相干断层扫描(SD - OCT)仪器测量视乳头周围视网膜神经纤维层(RNFL)厚度的一致性。
这是一项观察性横断面研究。使用Cirrus OCT(美国加利福尼亚州都柏林市卡尔·蔡司医疗技术公司)和Topcon OCT(3D - OCT 1000,日本东京拓普康公司)对89名受试者的双眼进行成像,其中44名健康受试者(平均年龄47±15岁),45名青光眼患者(平均年龄55±14岁)。在每个受试者中随机选择一只眼睛,并比较视乳头周围RNFL厚度测量值。采用Bland - Altman图评估一致性水平。
与Cirrus OCT相比,Topcon OCT测量的正常眼(101±12 vs 91±8 µm,p<0.001)和青光眼眼(89±18 vs 76±16 µm,p<0.001)的平均视乳头周围RNFL厚度明显更大。正常眼平均视乳头周围RNFL厚度测量的一致性界限较差(-11至32 µm),青光眼眼为(-4至30 µm)。
在正常眼和青光眼眼中,Topcon OCT获得的视乳头周围RNFL厚度值均显著高于Cirrus OCT测量的值。临床医生在监测由不同OCT仪器成像的患者时应注意这种差异;测量值不能互换使用。