Breivik K, Ohm O J, Engedal H
Pacing Clin Electrophysiol. 1982 Sep;5(5):650-7. doi: 10.1111/j.1540-8159.1982.tb02301.x.
Sixty-one patients given programmable pacemakers at initial implantation have been followed for a mean of 18 months (3-42) with non-invasive measurements of myocardial pulse-width threshold. Fifty of the patients had CPI 0505/0522 (Microlith-P/Microthin-PI) pacemakers with either Cordis 322-462 8 mm2 ball tip (n=12), Cordis 322-620 17.5 mm2 (n=23), or CPI 4116 porous tip electrodes (stimulation area 7.5 mm2 and sensing area 50 mm2) (n=15). Eleven patients had Medtronic 5985 (Spectrax-SX) pacemakers with either Medtronic 6907-R 8 mm2 ring tip (n=7), 6907 11 mm2 (n=3) or 6917 myocardial electrodes (area 12 mm2) (n=1). At acute implant, the ball tip and porous tip electrodes had the lowest stimulation thresholds, but the differences were only statistically significant in comparison with the 17.5 mm2 electrode (p less than 0.01). Chronically there were no significant differences between the various electrodes (p greater than 0.1), but the ball tip electrode tended to give best long-term results, and the porous tip electrode the poorest. Thirty-eight of the 61 patients (62.3%) had chronic pulse-width thresholds of 0.1 ms or less at approximately equal to 5 V output, indicating that pulse-width programming is a useful way to conserve battery energy. However, at some stage of the study, six of the patients (9.8%) had a pulse-width threshold of 0.5 ms or more. Pulse width should therefore not be set too narrow in standard nonprogrammable pacemakers.
61例初次植入可编程起搏器的患者接受了平均18个月(3 - 42个月)的随访,采用无创方法测量心肌脉冲宽度阈值。50例患者植入CPI 0505/0522(Microlith - P/Microthin - PI)起搏器,电极分别为Cordis 322 - 462 8平方毫米球头电极(n = 12)、Cordis 322 - 620 17.5平方毫米电极(n = 23)或CPI 4116多孔尖端电极(刺激面积7.5平方毫米,感知面积50平方毫米)(n = 15)。11例患者植入美敦力5985(Spectrax - SX)起搏器,电极分别为美敦力6907 - R 8平方毫米环形尖端电极(n = 7)、6907 11平方毫米电极(n = 3)或6917心肌电极(面积12平方毫米)(n = 1)。急性植入时,球头电极和多孔尖端电极的刺激阈值最低,但与17.5平方毫米电极相比,差异仅具有统计学意义(p小于0.01)。长期来看,各电极之间无显著差异(p大于0.1),但球头电极往往能给出最佳长期效果,而多孔尖端电极效果最差。61例患者中有38例(62.3%)在输出电压约为5V时慢性脉冲宽度阈值为0.1毫秒或更低,这表明脉冲宽度编程是节省电池能量的有效方法。然而,在研究的某个阶段,6例患者(9.8%)的脉冲宽度阈值为0.5毫秒或更高。因此,在标准的非可编程起搏器中,脉冲宽度不应设置得太窄。