Sündermann Simon H, Gessat Michael, Cesarovic Nikola, Frauenfelder Thomas, Biaggi Patric, Bettex Dominique, Falk Volkmar, Jacobs Stephan
Division of Cardiovascular Surgery, University Hospital Zurich, Zurich, Switzerland.
Interact Cardiovasc Thorac Surg. 2013 Apr;16(4):417-22. doi: 10.1093/icvts/ivs531. Epub 2013 Jan 3.
Implantation of an annuloplasty ring is an essential component of a durable mitral valve repair. Currently available off-the-shelf rings still do not cover all the variations in mitral annulus anatomy and pathology from subject to subject. Computed tomography (CT) and echo imaging allow for 3-D segmentation of the mitral valve and mitral annulus. The concept of tailored annuloplasty rings has been proposed although, to date, no surgically applicable implementation of patient-specific annuloplasty rings has been seen. The objective of this trial was to prove the concept of surgical implantation of a model-guided, personalized mitral annuloplasty ring, manufactured based on individual CT-scan models.
ECG-gated CT angiography was performed in six healthy pigs under general anaesthesia. Based on the individual shape of the mitral annulus in systole, a customized solid ring with integrated suturing holes was designed and manufactured from a biocompatible titanium alloy by a rapid process using laser melting. The ring was implanted three days later and valve function was assessed by intraoperative echocardiography. The macroscopic annulus-annuloplasty match was assessed after heart explantation.
CT angiography provided good enough image quality in all animals to allow for segmentation of the mitral annulus. The individually tailored mitral rings were manufactured and successfully implanted in all pigs. In 50%, a perfect matching of the implanted ring and the mitral annulus was achieved. In one animal, a slight deviation of the ring shape from the circumference was seen postoperatively. The rings implanted in the first two animals were significantly oversized but the deviation did not affect valve competence.
CT image quality and accuracy of the dimensions of the mitral annulus were sufficient for digital modelling and rapid manufacturing of mitral rings. Implantation of individually tailored annuloplasty rings is feasible.
植入瓣环成形环是持久二尖瓣修复的重要组成部分。目前市售的现成瓣环成形环仍无法涵盖个体间二尖瓣环解剖结构和病理情况的所有差异。计算机断层扫描(CT)和超声成像可实现二尖瓣和二尖瓣环的三维分割。尽管已经提出了定制瓣环成形环的概念,但迄今为止,尚未见到可用于手术的患者特异性瓣环成形环。本试验的目的是证明基于个体CT扫描模型制造的模型引导、个性化二尖瓣瓣环成形环的手术植入概念。
对6只健康猪在全身麻醉下进行心电图门控CT血管造影。根据二尖瓣环在收缩期的个体形状,设计并通过激光熔化快速工艺由生物相容性钛合金制造出带有集成缝合孔的定制实心环。三天后植入该环,并通过术中超声心动图评估瓣膜功能。心脏取出后评估宏观瓣环与瓣环成形环的匹配情况。
CT血管造影在所有动物中提供了足够好的图像质量,以允许对二尖瓣环进行分割。成功制造并将个体化定制的二尖瓣环植入所有猪体内。50%的猪实现了植入环与二尖瓣环的完美匹配。在一只动物中,术后发现环的形状与周长略有偏差。植入前两只动物体内的环明显过大,但这种偏差并未影响瓣膜功能。
CT图像质量和二尖瓣环尺寸的准确性足以用于二尖瓣环的数字建模和快速制造。植入个体化定制的瓣环成形环是可行的。