Suppr超能文献

CyberKnife 中电离室剂量学的校正因子:机器特异性、计划类别和临床领域。

Correction factors for ionization chamber dosimetry in CyberKnife: machine-specific, plan-class, and clinical fields.

机构信息

Universidade de Santiago de Compostela, A Coruña, Spain.

出版信息

Med Phys. 2013 Jan;40(1):011721. doi: 10.1118/1.4773047.

Abstract

PURPOSE

The aim of this work is the application of the formalism for ionization chamber reference dosimetry of small and nonstandard fields [R. Alfonso, P. Andreo, R. Capote, M. S. Huq, W. Kilby, P. Kjäll, T. R. Mackie, H. Palmans, K. Rosser, J. Seuntjens, W. Ullrich, and S. Vatnitsky, "A new formalism for reference dosimetry of small and nonstandard fields," Med. Phys. 35, 5179-5186 (2008)] to the CyberKnife robotic radiosurgery system. Correction factors for intermediate calibration fields, a machine-specific reference field (msr) and two plan-class specific reference fields (pcsr), have been studied. Furthermore, the applicability of the new formalism to clinical dosimetry has been analyzed through the investigation of two clinical treatments.

METHODS

PTW31014 and Scanditronix-Wellhofer CC13 ionization chamber measurements were performed for the fields under investigation. Absorbed dose to water was determined using alanine reference dosimetry, and experimental correction factors were calculated from alanine to ionization chamber readings ratios. In addition, correction factors were calculated for the intermediate calibration fields and one of the clinical treatment fields using the Monte Carlo method and these were compared with the experimental values.

RESULTS

Overall correction factors deviating from unity by approximately 2% were obtained from both measurements and simulations, with values below and above unity for the studied intermediate calibration fields and clinical fields for the ionization chambers under consideration. Monte Carlo simulations yielded correction factors comparable with those obtained from measurements for the machine-specific reference field, although differences from 1% to 3.3% were observed between measured and calculated correction factors for the composite intermediate calibration fields. Dose distribution inhomogeneities are thought to be responsible for such discrepancies.

CONCLUSIONS

The differences found between overall correction factors associated with the proposed intermediate calibration fields and the clinical fields under investigation show that depending on the clinical field and the detector used, either a machine-specific reference field or a plan-class specific reference field is more representative for the clinical field. Given the experimental and numerical uncertainties and the small number of clinical fields considered in this study the significance of these observations is limited and it remains unclear for the CyberKnife if there would be a significant gain in using a pcsr field rather than a msr field as reference field for relative dosimetry.

摘要

目的

本工作的目的是将小尺寸和非标准野的电离室参考剂量学公式[R. Alfonso、P. Andreo、R. Capote、M. S. Huq、W. Kilby、P. Kjäll、T. R. Mackie、H. Palmans、K. Rosser、J. Seuntjens、W. Ullrich 和 S. Vatnitsky,“小尺寸和非标准野参考剂量学的新公式”,《医学物理学》35,5179-5186(2008)]应用于 CyberKnife 机器人放射外科系统。研究了中间校准场、机器特定参考场(msr)和两个计划类别特定参考场(pcsr)的校正因子。此外,通过研究两种临床治疗方法,分析了新公式在临床剂量学中的适用性。

方法

对所研究的场进行了 PTW31014 和 Scanditronix-Wellhofer CC13 电离室测量。使用丙氨酸参考剂量学确定水吸收剂量,并通过从丙氨酸到电离室读数的比值计算实验校正因子。此外,使用蒙特卡罗方法计算了中间校准场和其中一个临床治疗场的校正因子,并将其与实验值进行了比较。

结果

从测量和模拟中得到的总校正因子与 1 的偏差约为 2%,对于所考虑的电离室,研究的中间校准场和临床场的校正因子低于和高于 1。蒙特卡罗模拟得到的校正因子与从测量中得到的机器特定参考场的校正因子相当,尽管对于复合中间校准场,实测和计算的校正因子之间存在 1%至 3.3%的差异。认为剂量分布不均匀是造成这种差异的原因。

结论

与所研究的拟议中间校准场和临床场相关的总校正因子之间的差异表明,取决于临床场和使用的探测器,机器特定参考场或计划类别特定参考场更能代表临床场。考虑到实验和数值不确定性以及本研究中考虑的临床场数量较少,这些观察结果的意义有限,并且对于 CyberKnife 是否使用 pcsr 场而不是 msr 场作为相对剂量学的参考场会有显著收益尚不清楚。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验