Suppr超能文献

青光眼患者视野平均偏差值的双眼间差异与自动瞳孔计测量的相对传入性瞳孔反应的相关性。

Correlation between intereye difference in visual field mean deviation values and relative afferent pupillary response as measured by an automated pupillometer in subjects with glaucoma.

机构信息

*Department of Ophthalmology, Northwestern University Feinberg School of Medicine, Chicago, IL †Konan Medical USA, Irvine, CA.

出版信息

J Glaucoma. 2014 Sep;23(7):419-23. doi: 10.1097/IJG.0b013e31827b1522.

Abstract

PURPOSE

To evaluate the effectiveness of a new binocular infrared computerized pupillometer in the quantitative measurement of the relative afferent pupillary response in patients with glaucoma by assessing the correlation of the intereye difference in visual function as measured by standard automated perimetry (SAP) with the intereye difference in the afferent pupillary response.

METHODS

Twenty-three patients with glaucoma underwent examination with a prototype, automated, binocular pupillometer. Correlation between the intereye difference in the afferent pupillary response and the intereye difference in mean deviation (MD) was explored.

RESULTS

Within 7 months of pupillography, all patients underwent SAP using the Humphrey Field Analyzer IIi, 24-2, Swedish Interactive Threshold Algorithm. The intereye differential pupillary response was 0.69±0.59 (log units, mean±SD). The intereye difference in MD was 5.67±5.29 dB (mean±SD). There was a strong correlation between the intereye difference in the afferent pupillary response and the intereye difference in MD (Spearman correlation coefficient, r = -0.77; P<0.001).

CONCLUSIONS

A new, binocular computerized pupillometer provides an automated method for the quantitative assessment of the afferent pupillary response. The intereye asymmetry in the pupil response correlates strongly with asymmetry in visual function, as measured by SAP, in patients with glaucoma.

摘要

目的

通过评估标准自动视野计(SAP)测量的视功能的眼间差异与传入性瞳孔反应的眼间差异之间的相关性,来评估新型双目红外电脑瞳孔计在定量测量青光眼患者相对传入性瞳孔反应中的有效性。

方法

对 23 例青光眼患者进行原型自动双目瞳孔计检查。探索传入性瞳孔反应的眼间差异与平均偏差(MD)的眼间差异之间的相关性。

结果

在瞳孔图检查后 7 个月内,所有患者均使用 Humphrey Field Analyzer IIi,24-2,瑞典互动阈值算法进行 SAP。传入性瞳孔反应的眼间差异为 0.69±0.59(对数单位,均值±标准差)。MD 的眼间差异为 5.67±5.29 dB(均值±标准差)。传入性瞳孔反应的眼间差异与 MD 的眼间差异之间存在很强的相关性(Spearman 相关系数 r = -0.77;P<0.001)。

结论

新型双目电脑瞳孔计提供了一种自动定量评估传入性瞳孔反应的方法。青光眼患者的瞳孔反应的眼间不对称与 SAP 测量的视功能的眼间不对称具有很强的相关性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验