Hopper Kate, Epstein Steven E
Department of Veterinary Surgical and Radiological Sciences, University of California, Davis, CA 95616, USA.
J Vet Emerg Crit Care (San Antonio). 2013 Jan-Feb;23(1):63-7. doi: 10.1111/j.1476-4431.2012.00825.x.
To describe false increases in plasma lactate concentration measured on point-of-care analyzers in 2 dogs with ethylene glycol (EG) intoxication.
Two dogs presenting with EG intoxication had extreme increases of plasma lactate concentrations recorded on a point-of-care machine. Laboratory analysis by spectrophotometry of lactate concentration determined these lactate measurements to be erroneous. False increases in plasma lactate concentration were demonstrated in 2 out of 3 point-of-care machines tested.
Glycolate, a toxic metabolite of EG, can interfere with the measurement of plasma lactate by some analyzers and this may delay the correct diagnosis of EG toxicity if not recognized.