Skogerboe K J, Labbé R F, Rettmer R L, Sundquist J P, Gargett A M
Department of Laboratory Medicine, University of Washington, Seattle 98195.
Clin Chem. 1990 May;36(5):752-5.
This instrumental method for total urinary nitrogen (TUN) is based on the principle of gas-phase chemiluminescence. Results correlate well with measurements of TUN by the Kjeldahl method, which has long provided the means to calculate nitrogen balances for nutritional management. In recent years, because of speed and convenience of measurement, determination of urinary urea nitrogen (UUN) has been substituted for Kjeldahl TUN. However, in patients requiring aggressive nutritional support, the UUN may not be a valid indicator of total nitrogen excretion. We compared nitrogen balances calculated for patients, using both UUN and chemiluminescence TUN data. For both normal and hospitalized populations, nitrogen balance calculated from UUN data exceeded that calculated from TUN data. We show that use of UUN data in calculating nitrogen balance may result in an incorrect assessment of many patients as being in positive nitrogen balance. TUN determined by chemiluminescence evidently provides a simple means of calculating nitrogen balance more nearly accurately.
这种测定总尿氮(TUN)的仪器方法基于气相化学发光原理。其结果与凯氏定氮法测定的TUN结果高度相关,长期以来凯氏定氮法一直是计算营养管理中氮平衡的方法。近年来,由于测定速度快且方便,尿尿素氮(UUN)的测定已取代凯氏定氮法测定的TUN。然而,对于需要积极营养支持的患者,UUN可能不是总氮排泄的有效指标。我们比较了使用UUN和化学发光法测定的TUN数据为患者计算的氮平衡。对于正常人群和住院患者,根据UUN数据计算的氮平衡超过了根据TUN数据计算的氮平衡。我们发现,在计算氮平衡时使用UUN数据可能会导致对许多患者处于正氮平衡的评估错误。通过化学发光法测定的TUN显然提供了一种更准确计算氮平衡的简单方法。