Gholizadeh Hossein, Abu Osman Noor Azuan, Eshraghi Arezoo, Ali Sadeeq, Sævarsson Stefan Karl, Wan Abas Wan Abu Bakar, Pirouzi Gholam Hossein
Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, Kuala Lumpur, Malaysia.
J Rehabil Res Dev. 2012;49(9):1321-30. doi: 10.1682/jrrd.2011.11.0221.
Poor suspension increases slippage of the residual limb inside the socket during ambulation. The main purpose of this article is to evaluate the pistoning at the prosthetic liner-socket interface during gait and assess patients' satisfaction with two different liners. Two prostheses with seal-in and locking liners were fabricated for each of the 10 subjects with transtibial amputation. The Vicon motion system was used to measure the pistoning during gait. The subjects were also asked to complete a Prosthesis Evaluation Questionnaire. The results revealed higher pistoning inside the socket during gait with the locking liner than with the seal-in liner (p < 0.05). The overall satisfaction with the locking liner was higher (p < 0.05) because of the relative ease with which the patients could don and doff the device. As such, pistoning may not be the main factor that determines patients' overall satisfaction with the prosthesis and other factors may also contribute to comfort and satisfaction with prostheses. The article also verifies the feasibility of the Vicon motion system for measuring pistoning during gait.
较差的悬吊性会增加步行过程中残肢在接受腔内的滑动。本文的主要目的是评估步态期间假肢内衬与接受腔界面处的活塞运动情况,并评估患者对两种不同内衬的满意度。为10名经胫骨截肢患者每人制作了带有密封式和锁定式内衬的两种假肢。使用Vicon运动系统测量步态期间的活塞运动。还要求受试者完成一份假肢评估问卷。结果显示,与密封式内衬相比,使用锁定式内衬时步态期间接受腔内的活塞运动更高(p < 0.05)。由于患者穿戴和脱下该装置相对容易,对锁定式内衬的总体满意度更高(p < 0.05)。因此,活塞运动可能不是决定患者对假肢总体满意度的主要因素,其他因素也可能影响对假肢的舒适度和满意度。本文还验证了Vicon运动系统在测量步态期间活塞运动方面的可行性。