Katnik C, Waugh R
Department of Biophysics, University of Rochester, School of Medicine and Dentistry, New York 14642.
Biophys J. 1990 Apr;57(4):877-82. doi: 10.1016/S0006-3495(90)82607-2.
Red blood cell membrane exhibits a large resistance to changes in surface area. This resistance is characterized by the area expansivity modulus K, which relates the isotropic membrane force resultant, T, to the fractional change in membrane surface area delta A/Ao. The experimental technique commonly used to determine K is micropipette aspiration. Using this method, E. A. Evans and R. Waugh (1977, Biophys. J. 20:307-313) obtained a value of 450 dyn/cm for the modulus. In the present report, it is shown that the value of K, as determined using this method, is affected by electric potential differences applied across the tip of the pipette. Using Ag-AgCl electrodes and current clamping electronics, we obtained values for K ranging from 150 dyn/cm with -1.0 V applied, to 1,500 dyn/cm with 1.0 V applied. At 0.0 V the modulus obtained was approximately 500 dyn/cm. A reversible, voltage- and pressure-dependent change in the cell volume probably accounts for the effect of the voltage on the calculated value of the modulus. The use of lanthanum chloride or increasing the extra- and intracellular solute concentrations reduced the voltage dependence of the measurements. It was also found that when dissimilar metals were used to "ground" the pipette to the chamber to prevent lysis of cells by static charge, values for K ranged from 121 to 608 dyn/cm. Based on measurements made at zero applied volts, in the presence of 0.4 mM lanthanum and at high solute concentration, we conclude that the true value of the modulus is approximately 500 dyn/cm.
红细胞膜对表面积变化表现出很大的抗性。这种抗性由面积膨胀模量K表征,它将各向同性的膜合力T与膜表面积的分数变化ΔA/A₀联系起来。常用于测定K的实验技术是微量吸管吸液法。利用这种方法,E. A. 埃文斯和R. 沃augh(1977年,《生物物理学杂志》20:307 - 313)得到模量的值为450达因/厘米。在本报告中,表明用这种方法测定的K值受施加在吸管尖端的电势差影响。使用银 - 氯化银电极和电流钳制电子设备,我们得到的K值范围从施加 - 1.0伏时的150达因/厘米到施加1.0伏时的1500达因/厘米。在0.0伏时得到的模量约为500达因/厘米。细胞体积的可逆的、电压和压力依赖性变化可能解释了电压对模量计算值的影响。使用氯化镧或增加细胞外和细胞内溶质浓度降低了测量的电压依赖性。还发现当使用不同金属将吸管“接地”到腔室以防止细胞因静电荷而裂解时,K值范围为121至608达因/厘米。基于在零施加电压、存在0.4毫摩尔氯化镧和高溶质浓度下进行的测量,我们得出模量的真实值约为500达因/厘米。