Suppr超能文献

比较传统侧向流动检测法和新的荧光免疫检测法检测流感病毒。

Comparison of conventional lateral-flow assays and a new fluorescent immunoassay to detect influenza viruses.

机构信息

Department of Pathology & Laboratories, Nassau University Medical Center, East Meadow, NY, USA.

出版信息

J Virol Methods. 2013 May;189(2):379-82. doi: 10.1016/j.jviromet.2013.02.008. Epub 2013 Feb 28.

Abstract

Sofia, a novel, fluorescent lateral-flow immunoassay was compared with two conventional colorimetric assays, Quickvue Influenza A+B and Directigen FLU A+B, to identify influenza viral antigen from patient nasopharyngeal specimens. A total of 118 frozen original influenza-positive specimens and 57 prospective specimens were examined. Using rt-PCR as a referee assay, sensitivity values (%) for influenza A/B of 80.0/74.8, 73.3/59.3 and 73.3/40.7 were obtained using the Sofia, Quickvue and Directigen assays, respectively. All assays demonstrated reduced sensitivity for influenza B as compared with influenza A virus. With respect to the Sofia assay, the sensitivity of influenza B for the Directigen assay was significantly diminished. False positive results were not observed in the Sofia and Directigen assays. The Quickvue assay produced 3 false-positive results (2 influenza A and 1 influenza B) resulting in a specificity (%) of 96 and 98 for influenza A and B, respectively. Cross-reactivity to other respiratory viruses was not observed among immunoassays. A sensitivity rank (highest to low) of rt-PCR>culture>Sofia>Quickvue>Directigen was established using dilutions of influenza A and B. Sofia provides enhanced sensitivity and objective result interpretation over conventional colorimetric immunoassays.

摘要

Sofia,一种新型荧光侧向流动免疫分析法,与两种传统比色分析法(Quickvue Influenza A+B 和 Directigen FLU A+B)进行比较,以鉴定患者鼻咽标本中的流感病毒抗原。共检测了 118 份冷冻原始流感阳性标本和 57 份前瞻性标本。以 rt-PCR 为参考检测方法,Sofia、Quickvue 和 Directigen 检测方法对流感 A/B 的灵敏度值(%)分别为 80.0/74.8、73.3/59.3 和 73.3/40.7。与流感 A 病毒相比,所有检测方法均显示流感 B 的灵敏度降低。与 Sofia 检测法相比,Directigen 检测法中流感 B 的灵敏度显著降低。在 Sofia 和 Directigen 检测法中未观察到假阳性结果。Quickvue 检测法产生了 3 个假阳性结果(2 个流感 A 和 1 个流感 B),导致流感 A 和 B 的特异性(%)分别为 96 和 98。免疫检测法之间未观察到与其他呼吸道病毒的交叉反应。使用流感 A 和 B 的稀释液建立了灵敏度等级(从高到低):rt-PCR>培养>Sofia>Quickvue>Directigen。Sofia 提供了比传统比色免疫分析法更高的灵敏度和客观的结果解释。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验