With little data published on the accuracy of cardiac 3-D strain measurements, we investigated the agreement between 3-D echocardiography and sonomicrometry in an in vitro model with a polyvinyl alcohol phantom. A cardiac scanner with a 3-D probe was used to acquire recordings at 15 different stroke volumes at a heart rate of 60 beats/min, and eight different stroke volumes at a heart rate of 120 beats/min. Sonomicrometry was used as a reference, monitoring longitudinal, circumferential and radial lengths. Both single- and multi-beat acquisitions were recorded. Strain values were compared with sonomicrometer strain using linear correlation coefficients and Bland-Altman analysis. Multi-beat acquisition showed good agreement, whereas real-time images showed less agreement. The best correlation was obtained for a heart rate 60 of beats/min at a volume rate 36.6 volumes/s.
由于关于心脏 3-D 应变测量准确性的公开数据较少,我们使用含聚乙烯醇的水凝胶模型研究了 3-D 超声心动图与超声心动描记术之间的体外模型的一致性。使用具有 3-D 探头的心脏扫描仪,在心率为 60 次/分钟时,以 15 个不同的每搏量获取记录,在心率为 120 次/分钟时,以 8 个不同的每搏量获取记录。超声心动描记术作为参考,监测纵向、周向和径向长度。记录了单次和多次心跳获取。使用线性相关系数和 Bland-Altman 分析比较应变值与超声心动描记术应变。多心跳获取显示出很好的一致性,而实时图像显示出较小的一致性。在体积率为 36.6 毫升/秒、心率为 60 次/分钟时,相关性最好。