Department of Radiology, Seoul National University Hospital, Seoul, Korea.
J Magn Reson Imaging. 2014 Feb;39(2):326-31. doi: 10.1002/jmri.24147. Epub 2013 Apr 15.
To determine the reproducibility of MR elastography (MRE) and the reproducibility and repeatability of the stiffness measurement of MRE in the staging of liver fibrosis.
Ninety-four patients, who underwent liver MRE, were included in this study. The patients were classified into group 1 (n = 47) and group 2 (n = 47) according to our knowledge of their histologic hepatic fibrosis (HF) stage. To analyze the reproducibility of MRE, the group 1 patients underwent MRE twice. In addition, to evaluate the repeatability and reproducibility of the stiffness measurement of MRE, a single observer measured the stiffness values of the second MREs in group 1 twice, and two observers independently measured the stiffness values of MRE in group 2. A 95% Bland-Altman limits of agreement and intraclass correlation coefficients (ICCs) were used to evaluate the reproducibility and repeatability of MRE.
In group I, there was no significant difference in the mean liver stiffness values of the first and second MRE examinations, i.e., 3.45 ± 0.25 kPa vs. 3.35 ± 0.23 kPa (p = 0.22). The reproducibility of the MRE examination and the reproducibility and repeatability of the stiffness measurement were high, i.e., the ICCs of each parameter were 0.945, 0.827, and 0.963, respectively, and the 95% limits of agreement were 25.3%, 35.35%, and 18.0%, respectively.
MRE is a promising tool for evaluating HF and has high reproducibility of the examination as well as reproducibility and repeatability of the stiffness measurements.
确定磁共振弹性成像(MRE)的可重复性,以及 MRE 硬度测量在肝纤维化分期中的可重复性和再现性。
本研究纳入 94 例行肝脏 MRE 的患者。根据我们对其组织学肝纤维化(HF)分期的了解,将患者分为组 1(n=47)和组 2(n=47)。为了分析 MRE 的可重复性,组 1 患者进行了两次 MRE。此外,为了评估 MRE 硬度测量的可重复性和再现性,一位观察者对组 1 的第二次 MRE 硬度值进行了两次测量,两位观察者独立测量了组 2 的 MRE 硬度值。使用 95% Bland-Altman 一致性界限和组内相关系数(ICC)评估 MRE 的可重复性和再现性。
在组 1 中,首次和第二次 MRE 检查的肝脏硬度平均值无显著差异,分别为 3.45±0.25 kPa 与 3.35±0.23 kPa(p=0.22)。MRE 检查的可重复性以及硬度测量的可重复性和再现性均较高,即各参数的 ICC 分别为 0.945、0.827 和 0.963,95%一致性界限分别为 25.3%、35.35%和 18.0%。
MRE 是评估 HF 的一种很有前途的工具,具有较高的检查可重复性以及硬度测量的可重复性和再现性。