健康肝实质的磁共振弹性成像:肝硬度值测量的正常值和可靠性。

MR elastography of healthy liver parenchyma: Normal value and reliability of the liver stiffness value measurement.

机构信息

Department of Radiology, Suncheon OH Hospital, Suncheon City, Korea.

出版信息

J Magn Reson Imaging. 2013 Nov;38(5):1215-23. doi: 10.1002/jmri.23958. Epub 2012 Dec 19.

Abstract

PURPOSE

To investigate the normal liver stiffness value using MR elastography (MRE) and to compare the repeatability and reproducibility of three measurement methods.

MATERIALS AND METHODS

Forty-nine subjects who underwent liver MR imaging including elastography using a 1.5 Tesla scanner, who had normal laboratory test results, and who underwent liver donation, were included in this study. Two, independent readers measured the liver stiffness value of the hepatic parenchyma using three methods, including evaluating various liver parenchyma portions using different-sized, regions-of-interest (ROIs): 2cm-ROI-per-slice in the right lobe (2cm-per-slice); 1cm-ROI-per-segment (1cm-S); and 70%-of-the-surface area (70% S). The mean liver stiffness values were compared between gender using the t-test and among the age groups using one-way analysis of variance. The reproducibility and repeatability of the three liver stiffness value measurement methods were determined using intraclass correlation coefficients (ICCs) and the 95% Bland-Altman limits of agreement.

RESULTS

The liver stiffness values in living liver donors ranged from 1.54 to 2.87 kPa. The mean stiffness value was 2.05 kilopascal using the 2cm-per-slice method, 2.01 kilopascal with the 1cm-S method and 2.12 kilopascal using the 70% S method. There was no significant difference in the liver stiffness value according to the gender or age factors. For reproducibility, the ICCs were 0.416 with the 2cm-per-slice method, 0.800 using the 1cm-S method, and 0.845 with the 70% S method. In terms of repeatability, the ICCs were 0.238 using the 2cm-per-slice method, 0.914 with the 1cm-S method, and 0.852 using the 70% S method. The ICCs determined using the 2cm-per-slice method were significantly lower than those of the 1cm-S or 70% S method for assessing both reproducibility and repeatability. The 95% limit of agreement was 54.0% with the 2cm-per-slice method, 24.0% using the 1cm-S method, and 18.8% with the 70% S method, in terms of reproducibility. To assess the repeatability, the 95% limit of agreement was 63.3% using the 2cm-per-slice method, 20.6% with the 1cm-S method, and 17.4% using the 70% S method.

CONCLUSION

The mean liver stiffness values in living donors ranged from 2.05 to 2.12 kilopascals and did not differ significantly for either gender or age. The 1cm-S and 70% S methods were significantly more reliable compared with the 2cm-per-slice method.

摘要

目的

利用磁共振弹性成像(MRE)研究正常肝脏硬度值,并比较三种测量方法的可重复性和再现性。

材料和方法

本研究纳入了 49 名在 1.5T 扫描仪上进行肝脏 MRI 检查(包括弹性成像)的志愿者,这些志愿者实验室检查结果正常,并计划捐献肝脏。两名独立的读者使用三种方法测量肝实质的肝脏硬度值,包括使用不同大小的感兴趣区(ROI)评估各种肝实质部分:右叶 2cm-ROI/层(2cm-per-slice);1cm-ROI/段(1cm-S);70%的表面积(70% S)。使用 t 检验比较性别间的平均肝脏硬度值,使用单因素方差分析比较年龄组间的平均肝脏硬度值。使用组内相关系数(ICC)和 95% Bland-Altman 一致性界限评估三种肝脏硬度值测量方法的可重复性和再现性。

结果

活体肝供者的肝脏硬度值范围为 1.54 至 2.87kPa。2cm-per-slice 方法的平均硬度值为 2.05kPa,1cm-S 方法为 2.01kPa,70% S 方法为 2.12kPa。肝脏硬度值不受性别或年龄因素的影响。在可重复性方面,2cm-per-slice 方法的 ICC 为 0.416,1cm-S 方法为 0.800,70% S 方法为 0.845。在再现性方面,2cm-per-slice 方法的 ICC 为 0.238,1cm-S 方法为 0.914,70% S 方法为 0.852。2cm-per-slice 方法的 ICC 明显低于 1cm-S 或 70% S 方法,无论是评估可重复性还是再现性。在可重复性方面,2cm-per-slice 方法的 95%一致性界限为 54.0%,1cm-S 方法为 24.0%,70% S 方法为 18.8%。在评估再现性方面,2cm-per-slice 方法的 95%一致性界限为 63.3%,1cm-S 方法为 20.6%,70% S 方法为 17.4%。

结论

活体供者的平均肝脏硬度值范围为 2.05 至 2.12kPa,性别或年龄对其无显著影响。1cm-S 和 70% S 方法比 2cm-per-slice 方法更可靠。