Rohlmann Antonius, Zander Thomas, Graichen Friedmar, Bergmann Georg
Julius Wolff Institute, Charité-Universitätsmedizin Berlin, Augustenburger Platz 1, 13353 Berlin, Germany.
Clin Biomech (Bristol). 2013 Jun;28(5):490-4. doi: 10.1016/j.clinbiomech.2013.03.010. Epub 2013 Apr 23.
The spinal load reduction by an orthosis is still a matter of debate. Some studies predicted a load reduction while others found no effect. The aim of this study was to measure the in vivo effect of the Lumbo TriStep brace and the hyperextension orthosis medi 3C on the spinal implant loads.
Telemeterized vertebral body replacements were implanted in 5 patients suffering from a severe fracture of the L1 or L3 vertebral body. The implant allows the measurement of 6 load components acting on it. For several activities during standing, sitting and walking, implant loads were measured in patients with and without an orthosis.
The average resultant force on the vertebral body for 26 activities was reduced by 9% with the Lumbo TriStep brace, and by 19% with the hyperextension orthosis. The force reduction is usually more pronounced for activities performed during sitting than it is for those performed while standing. However, considerable inter- and intra-individual variation was observed. In several cases, the measured implant forces were even higher when the patients were wearing an orthosis.
In some patients, for certain activities, an orthosis may reduce the force on a vertebral body replacement and thus on the anterior column of the spine. However, in other patients for the same activities, an orthosis may increase the force. The measurements do not allow a clear recommendation to wear an orthosis since the clinically relevant reduction of implant forces is unknown.
矫形器对脊柱负荷的减轻作用仍存在争议。一些研究预测负荷会减轻,而另一些研究则未发现效果。本研究的目的是测量Lumbo TriStep支具和超伸展矫形器medi 3C对脊柱植入物负荷的体内影响。
将遥测椎体置换物植入5例患有L1或L3椎体严重骨折的患者体内。该植入物可测量作用于其上的6个负荷分量。在站立、坐立和行走期间的几种活动中,测量佩戴和不佩戴矫形器患者的植入物负荷。
使用Lumbo TriStep支具时,26种活动中椎体上的平均合力降低了9%,使用超伸展矫形器时降低了19%。坐立期间进行的活动中,力的降低通常比站立时进行的活动更明显。然而,观察到个体间和个体内存在相当大的差异。在某些情况下,患者佩戴矫形器时测量到的植入物力甚至更高。
在某些患者中,对于某些活动,矫形器可能会降低椎体置换物上的力,从而降低脊柱前柱上的力。然而,在其他患者中,对于相同的活动,矫形器可能会增加力。由于植入物力在临床上的相关降低情况未知,这些测量结果无法明确建议佩戴矫形器。