解剖型柄的缩短,多短才算足够短?一项关于负荷传递和初始稳定性的体外研究。

Shortening of an anatomical stem, how short is short enough? An in vitro study of load transfer and primary stability.

作者信息

Østbyhaug Per Olav, Klaksvik Jomar, Romundstad Pål, Aamodt Arild

机构信息

Department of Orthopaedic Surgery, Trondheim University Hospital, Trondheim, Norway.

出版信息

Proc Inst Mech Eng H. 2013 May;227(5):481-9. doi: 10.1177/0954411912471779. Epub 2013 Feb 5.

Abstract

An anatomical stem should be short enough to avoid distal locking and distal load transfer but long enough to ensure adequate primary stability of the stem. In this in vitro study, the cortical strains in the femur and the primary stability of the stem were measured after insertion of Anatomic Benoist Girard-I anatomical stems with gradually reduced stem length in six human cadaver femurs in order to find the optimal stem length. A shortening of 40-50 mm, corresponding to a stem extending 30-40 mm below the lesser trochanter, did not affect stem stability but nearly normalized the load distribution in the lower metaphysis and upper diaphysis. The large strain shielding observed in the calcar region was not influenced by shortening of the stem.

摘要

解剖型柄应足够短以避免远端锁定和远端负荷传递,但又要足够长以确保柄的足够初始稳定性。在这项体外研究中,为了找到最佳柄长度,在六具人类尸体股骨中插入柄长度逐渐缩短的解剖型贝努瓦·吉拉尔 - I型解剖柄后,测量了股骨中的皮质应变和柄的初始稳定性。缩短40 - 50毫米,相当于柄在小转子下方延伸30 - 40毫米,并不影响柄的稳定性,但几乎使干骺端下部和骨干上部的负荷分布正常化。在股骨矩区域观察到的大应变屏蔽不受柄缩短的影响。