Villarinho Eduardo Aydos, Cervieri André, Shinkai Rosemary Sadami Arai, Grossi Márcio Lima, Teixeira Eduardo Rolim
1 Private practice, Porto Alegre, RS, Brazil.
J Oral Implantol. 2015 Apr;41(2):139-43. doi: 10.1563/AAID-JOI-D-12-00245. Epub 2013 May 3.
The biomechanical stability of the implant-abutment connection is critical for the success of implant-supported restorations. This study investigated the effect of a positioning index on the abutment screw preload values of tapered connection implants. Twenty Morse taper implants presenting an internal locking hex received 10 solid and 10 straight screw retained abutments for cemented single-crown restorations. Ten abutments had a positioning index to fit the internal locking hex of the implant (straight), and 10 were locked only by the implant taper (solid). The preload values for each abutment screw after a tightening torque were registered by strain gauges. Prosthetic crowns were placed on each abutment and subjected to mechanical cycling. Detorque forces were applied to each abutment and compared with the initial torque values. Data were statistically analyzed using Kolmogorov-Smirnov and Student t tests. The nonindexed group presented higher initial preload (6.05 N ± 0.95 N) compared with the indexed group (4.88 N ± 0.92 N; P < .05). After cycling, the nonindexed group exhibited less reduction of preload (13.84% ± 6.43%) compared with the indexed group (52.65% ± 14.81%; P < .01). Indexed tapered abutments for single-crown restorations might represent greater biomechanical risk under function.
种植体-基台连接的生物力学稳定性对于种植体支持修复体的成功至关重要。本研究调查了定位指数对锥形连接种植体基台螺钉预紧力值的影响。二十个带有内部锁定六角的莫氏锥度种植体,用于粘结单冠修复,分别接收10个实心和10个直形螺钉固位基台。10个基台有一个定位指数以适配种植体的内部锁定六角(直形),10个仅通过种植体锥度锁定(实心)。拧紧扭矩后,通过应变片记录每个基台螺钉的预紧力值。在每个基台上放置修复冠并进行机械循环。对每个基台施加拧松力,并与初始扭矩值进行比较。使用柯尔莫哥洛夫-斯米尔诺夫检验和学生t检验对数据进行统计学分析。与有定位指数组(4.88 N±0.92 N;P<.05)相比,无定位指数组呈现出更高的初始预紧力(6.05 N±0.95 N)。循环后,与有定位指数组(52.65%±14.81%;P<.01)相比,无定位指数组的预紧力降低较少(13.84%±6.43%)。用于单冠修复的有定位指数的锥形基台在功能状态下可能代表更大的生物力学风险。