在机械循环中,基台角度对修复体基台扭矩损失的影响。

Influence of abutment angulation on loss of prosthetic abutment torque under mechanical cycling.

机构信息

Post Graduate student, Division of Oral Implantology, Faculdade São Leopoldo Mandic (SLMandic), Campinas, São Paulo, Brazil.

Professor, Division of Oral Implantology, Faculdade São Leopoldo Mandic (SLMandic), Campinas, São Paulo, Brazil.

出版信息

J Prosthet Dent. 2021 Feb;125(2):349.e1-349.e6. doi: 10.1016/j.prosdent.2020.10.010. Epub 2020 Nov 14.

Abstract

STATEMENT OF PROBLEM

Internal conical connections provide mechanical stability for the prosthetic abutment and implant connection. However, some clinical situations require the use of angled prosthetic abutments that may increase stress on supportive implants by difference force vectors under cyclic loading.

PURPOSE

The purpose of this in vitro study was to measure the screw loosening values of prosthetic abutments with internal conical connections (indexed and nonindexed) having different angles under mechanical cycling.

MATERIAL AND METHODS

Thirty-six implants (4.0×13 mm, Titamax) with internal conical connections and their respective universal prosthetic abutments (n=36, 3.5×3.3 mm) were divided into indexed and nonindexed groups (n=18) with abutment inclinations of 0 (straight), 17, and 30 degrees. An insertion torque of 15 Ncm was applied according to the manufacturer's specifications. The specimens underwent fatigue testing of 500 000 cycles at a frequency of 2 Hz with a dynamic compressive load of 120 N at an angle of 30 degrees. The detorque values were measured by using a digital torque meter and tabulated for statistical analyses.

RESULTS

The specimens with indexed abutments had mean ±standard deviation detorque values of 6.72 ±2.29 Ncm under mechanical cycling, whereas those with nonindexed abutments had values of 8.98 ±1.84 Ncm. In the indexed group, the lowest detorque value was observed for abutments at 30 degrees compared with the straight group (P<.05). As for nonindexed abutments, similar detorque values were observed after increasing the abutment inclination (P>.05).

CONCLUSIONS

A decrease in detorque values in the indexed abutments related to their inclination was found under mechanical cycling, whereas the prosthetic abutments with 30 degrees of angulation had the lowest values. No decrease was found in the nonindexed abutments.

摘要

问题陈述

内部锥形连接为修复体基台和种植体连接提供机械稳定性。然而,在某些临床情况下,需要使用角度式修复体基台,这可能会通过循环加载下的不同力向量增加对支持种植体的应力。

目的

本体外研究的目的是测量在机械循环下具有不同角度的内部锥形连接(带和不带角度)的修复体基台的螺钉松动值。

材料和方法

将 36 个带有内部锥形连接的种植体(4.0×13mm,Titamax)及其各自的通用修复体基台(n=36,3.5×3.3mm)分为带角度和不带角度的两组(n=18),基台倾斜度分别为 0(直)、17 和 30 度。根据制造商的规格施加 15Ncm 的插入扭矩。在频率为 2Hz、角度为 30 度的情况下,对标本进行 50 万次疲劳测试,施加动态压缩载荷为 120N。使用数字扭矩计测量脱扭矩值,并进行统计分析。

结果

在机械循环下,带有角度基台的标本的平均脱扭矩值为 6.72±2.29Ncm,而不带角度基台的标本的平均脱扭矩值为 8.98±1.84Ncm。在带角度组中,与直基台相比,30 度基台的脱扭矩值最低(P<.05)。对于不带角度的基台,随着基台倾斜度的增加,观察到相似的脱扭矩值(P>.05)。

结论

在机械循环下,发现角度式基台的脱扭矩值与其倾斜度有关,而 30 度角的修复体基台的脱扭矩值最低。非角度式基台未发现脱扭矩值降低。