7T 下颅外神经外科植入物的射频相关加热评估。

RF-related heating assessment of extracranial neurosurgical implants at 7T.

机构信息

The Ohio State University, Columbus, OH, USA.

出版信息

Magn Reson Imaging. 2013 Jul;31(6):1029-34. doi: 10.1016/j.mri.2012.10.025. Epub 2013 Apr 30.

Abstract

PURPOSE

The purpose was to evaluate radiofrequency (RF)-related heating of commonly used extracranial neurosurgical implants in 7-T magnetic resonance imaging (MRI).

MATERIALS AND METHODS

Experiments were performed using a 7-T MR system equipped with a transmit/receive RF head coil. Four commonly used titanium neurosurgical implants were studied using a test procedure adapted from the American Society for Testing and Materials Standard F2182-11a. Implants (n=4) were tested with an MRI turbo spin echo pulse sequence designed to achieve maximum RF exposure [specific absorption rate (SAR) level=9.9W/kg], which was further validated by performing calorimetry. Maximum temperature increases near each implant's surface were measured using fiberoptic temperature probes in a gelled-saline-filled phantom that mimicked the conductive properties of soft tissue. Measurement results were compared to literature data for patient safety.

RESULTS

The highest achievable phantom averaged SAR was determined by calorimetry to be 2.0±0.1W/kg due to the highly conservative SAR estimation model used by this 7-T MR system. The maximum temperature increase at this SAR level was below 1.0°C for all extracranial neurosurgical implants that underwent testing.

CONCLUSION

The findings indicated that RF-related heating under the conditions used in this investigation is not a significant safety concern for patients with the particular extracranial neurosurgical implants evaluated in this study.

摘要

目的

评估 7T 磁共振成像(MRI)中常用的颅外神经外科植入物的射频(RF)相关加热情况。

材料和方法

使用配备发射/接收 RF 头部线圈的 7T MR 系统进行实验。使用改编自美国测试材料协会标准 F2182-11a 的测试程序研究了四种常用的钛神经外科植入物。使用设计为实现最大 RF 暴露(比吸收率[SAR]水平=9.9W/kg)的 MRI 涡轮自旋回波脉冲序列对植入物(n=4)进行了测试,该测试通过进行量热法进一步验证。使用光纤温度探头在凝胶化盐水填充的仿体中测量每个植入物表面附近的最大温度升高,该仿体模拟软组织的导电特性。将测量结果与患者安全的文献数据进行比较。

结果

由于该 7T MR 系统使用的 SAR 估计模型非常保守,通过量热法确定的最高可达幻影平均 SAR 为 2.0±0.1W/kg。在所有接受测试的颅外神经外科植入物中,在该 SAR 水平下,最大温度升高低于 1.0°C。

结论

研究结果表明,在本研究中评估的特定颅外神经外科植入物的条件下,RF 相关加热不会对患者的安全性产生重大影响。