Missouri Department of Health and Senior Services, Jefferson City, Missouri 63103, USA.
Emerg Infect Dis. 2013 May;19(5):781-3. doi: 10.3201/eid1905.12039.
We investigated Bacillus cereus-positive tracheal aspirates from infants on ventilators in a neonatal intensive care unit. Multilocus sequence typing determined a genetic match between strains isolated from samples from a case-patient and from the air flow sensor in the ventilator. Changing the sterilization method for sensors to steam autoclaving stopped transmission.