Croft P R, Cruickshank J K
Department of Postgraduate Medicine, University of Keele, Stoke-on-Trent, Staffordshire, United Kingdom.
J Epidemiol Community Health. 1990 Jun;44(2):170-3. doi: 10.1136/jech.44.2.170.
The aim of the study was to determine whether a single size of cuff for adult blood pressure measurements is appropriate for general clinical practice.
The study was a prospective survey of a sample of adult blood pressure measurements using two cuffs with different bladder sizes (12 X 23 cm and 15 X 33 cm) in a randomised design using a random zero sphygmomanometer.
Blood pressures were measured in a general practice and in a hospital outpatient clinic.
The participants were 35-60 year old men and women invited to attend a blood pressure screening programme in the general practice (n = 170), and 35-74 year old patients attending a general medical outpatients (n = 72).
The small cuff gave higher readings of systolic blood pressure than the large cuff (mean difference 4.4 mm Hg). The difference increased as systolic pressure increased but did not show a clear association with arm circumference. The small cuff also gave higher diastolic pressure readings (mean difference 3.0 mm Hg), but only when arm circumference exceeded 30 cm. The variability of the differences between readings from the two cuffs was wide, little affected by arm circumference, and was similar to the variability between measurements using the same cuff size.
In terms of precision there is no basis for using two different cuff sizes unless it is physically difficult to obtain a reading with one or the other. Since readings with large cuffs are closer to intraarterial pressures in large arms, and the large cuff used here did not underestimate diastolic pressure in small arms, the large cuff alone could be recommended for general use.
本研究旨在确定用于成人血压测量的单一尺寸袖带是否适用于一般临床实践。
本研究采用随机零位血压计,以随机设计的方式,对使用两种不同气囊尺寸(12×23厘米和15×33厘米)袖带的成人群体进行血压测量的前瞻性调查。
在一家普通诊所和一家医院门诊进行血压测量。
参与者为受邀参加普通诊所血压筛查项目的35至60岁男性和女性(n = 170),以及普通内科门诊的35至74岁患者(n = 72)。
小袖带测得的收缩压读数高于大袖带(平均差值4.4毫米汞柱)。随着收缩压升高,差值增大,但与臂围无明显关联。小袖带测得的舒张压读数也较高(平均差值3.0毫米汞柱),但仅在臂围超过30厘米时出现。两种袖带读数差异的变异性较大,受臂围影响较小,且与使用相同袖带尺寸测量之间的变异性相似。
就精度而言,除非使用某一种袖带实际难以获得读数,否则没有理由使用两种不同尺寸的袖带。由于大袖带测得的读数更接近大臂的动脉内压力,且此处使用的大袖带并未低估小臂的舒张压,因此可推荐单独使用大袖带。