Fuglsang-Frederiksen A, Rønager J
Department of Clinical Neurophysiology, Hvidovre Hospital, University of Copenhagen, Denmark.
J Neurol Sci. 1990 Jun;97(1):81-91. doi: 10.1016/0022-510x(90)90100-2.
The diagnostic value of power spectrum analysis of the needle EMG pattern at a force of 30% of maximum was compared to that of turns-amplitude analysis and to that of manual measurements of motor unit potential (MUP) duration in the brachial biceps muscle of 20 patients with myopathy and 11 patients with neurogenic disorders. In myopathy the power spectrum analysis had the same diagnostic value as the turns-amplitude analysis and MUP duration measurements and the 3 methods supplemented each other. In patients with neurogenic disorders the diagnostic value of the power spectrum analysis as well as that of the turns-amplitude analysis were lesser than that of MUP duration measurement. In diseased muscles the amount of high frequencies increased with increasing ratio of turns to mean amplitude while there was no relation between the power spectrum and the MUP changes. The results suggest that the power spectrum analysis of EMG can be used as a diagnostic tool in patients with neuromuscular disorders.
将20例肌病患者和11例神经源性疾病患者肱二头肌在最大肌力30%时针电极肌电图模式的功率谱分析诊断价值,与转折-波幅分析以及运动单位电位(MUP)时限的手工测量诊断价值进行比较。在肌病中,功率谱分析与转折-波幅分析及MUP时限测量具有相同的诊断价值,且这三种方法相互补充。在神经源性疾病患者中,功率谱分析以及转折-波幅分析的诊断价值均低于MUP时限测量。在患病肌肉中,高频成分的量随转折与平均波幅比值的增加而增加,而功率谱与MUP变化之间无关联。结果提示,肌电图的功率谱分析可作为神经肌肉疾病患者的一种诊断工具。