Lieberman A
Particle Measuring Systems, Inc., Fremont, California 94578.
J Parenter Sci Technol. 1990 May-Jun;44(3):133-8.
Operation of an optical particle counter with microprocessor control of particle size calibration and data recording is described. Sensors are available to measure particles 0.1 micron and larger. A system with a range from 2 microns to 125 microns will be described. Calibration data based on response to monosized particles are stored in the system memory, along with sensor resolution information and system changes since the last calibration. The instrument adjusts the particle size channels internally for the selected calibration base particles. Once calibrated, the instrument can be set to report data in operator selectable sizes or any of four default particle size groups. Three default groups are based on pharmaceutical and hydraulic fluid power industry standards and one is set in a logarithmic spread of 16 particle sizes that covers the system dynamic range. A front-panel keyboard and CRT display allow selection and control of all operating parameters. These include reported size ranges, liquid sample size, number of sample replications (including suspect sample rejection), data averaging, and comparison of data from two sensors. Differential or cumulative data can be reported in tabular or histogram format. Data can be sent to an internal thermal printer or to an RS-232C output.
本文描述了一种具有微处理器控制粒度校准和数据记录功能的光学粒子计数器的操作。该计数器配备有可测量粒径为0.1微米及更大粒径颗粒的传感器。本文将介绍一种粒径范围为2微米至125微米的系统。基于对单分散颗粒响应的校准数据与传感器分辨率信息以及自上次校准以来的系统变化数据一起存储在系统内存中。该仪器会根据选定的校准基准颗粒在内部调整粒径通道。一旦校准完成,该仪器可设置为按操作员选择的粒径或四个默认粒径组中的任何一个来报告数据。其中三个默认组基于制药和液压流体动力行业标准,另一个组设置为16个粒径的对数分布,覆盖系统动态范围。前面板键盘和阴极射线管显示器允许选择和控制所有操作参数。这些参数包括报告的粒径范围、液体样品大小、样品重复次数(包括可疑样品剔除)、数据平均以及来自两个传感器的数据比较。差分或累积数据可以以表格或直方图格式报告。数据可以发送到内部热敏打印机或RS - 232C输出接口。