Gratama J W, Kraan J, Adriaansen H, Hooibrink B, Levering W, Reinders P, Van den Beemd M W, Van der Holt B, Bolhuis R L
Department of Clinical and Tumor Immunology, Daniel den Hoed Cancer Center, Rotterdam, The Netherlands.
Cytometry. 1997 Feb 15;30(1):10-22.
Two workshops addressed the question to which degree standardization of instrument set-up and calibration, and standard list mode data analysis would reduce interlaboratory variability of flow cytometric results on prestained peripheral blood mononuclear cells (PBMC). Standard instrument set-up included uniform positioning of the "windows of analysis" for the forward and sideward light scatter and fluorescence (FL) 1 (i.e., fluorescein isothiocyanate [FITC]) and 2 (i.e., phycoerythrin [PE]) parameters. Reference standards and PBMC, double-stained with FITC- and PE-conjugated monoclonal antibodies covering a wide range of FL intensities and coexpression patterns, were sent out to 25 laboratories in Workshop 1 and to 35 laboratories in Workshop 2 with the following requests: a) to set up instruments according to local and standard protocols, b) to acquire list mode data on the PBMC with both instrument settings, and c) to analyze both datasets according to local protocols. Standard analysis of the list mode data acquired with uniform instrument settings was performed centrally using so-called "latent class model" software (Van Putten et al., Cytometry 14:86-96, 1993). This software provides an automated, "no-gating" analytical method of lymphocyte immunophenotypes and employs fixed FL marker settings as defined prior to each analytical run. In Workshop 1, these markers were set in identical histogram channels for all instruments based on results obtained with a reference instrument. Standard analysis of list mode data acquired after uniform instrument set-up led only to a 13% reduction of interlaboratory variability of results as compared to data analysis using local protocols. The standard protocol for instrument set-up led to uniform positioning of relatively strong FL signals but variable positioning of unstained cells on the FL histogram scales. Hence, standard FL marker settings were inappropriate for some instruments. Therefore, instrument responses to FITC and PE signals in Workshop 2 were calibrated using microbeads labeled with FITC or PE in a range of predefined FL intensities expressed in MESF units (molecules of equivalent soluble fluorochrome). That approach allowed the positioning of the FL markers for the standard analysis on the basis of identical FL1 and FL2 intensities, expressed in MESF units, for all instruments. Standard analysis of list mode data acquired after uniform instrument set-up and calibrated FL marker settings led to a 43% reduction of interlaboratory variability as compared to data analysis to local protocols. We conclude that standard list mode data analysis using fixed FL marker settings reduces the interlaboratory variability of flow cytometric results on prestained PBMC, provided that the instruments have been set up in a uniform way and that FL markers have been standardized on the basis of calibration of each instrument's response to the corresponding FL signals.
两场研讨会探讨了仪器设置与校准的标准化以及标准列表模式数据分析在何种程度上能够降低流式细胞术检测预染色外周血单个核细胞(PBMC)结果的实验室间差异。标准仪器设置包括统一设定前向和侧向光散射以及荧光(FL)1(即异硫氰酸荧光素[FITC])和2(即藻红蛋白[PE])参数的“分析窗口”。参考标准品和用覆盖广泛FL强度及共表达模式的FITC和PE偶联单克隆抗体进行双重染色的PBMC被分发给研讨会1中的25个实验室以及研讨会2中的35个实验室,并提出以下要求:a)根据当地和标准方案设置仪器;b)使用两种仪器设置获取PBMC的列表模式数据;c)根据当地方案分析这两个数据集。使用统一仪器设置获取的列表模式数据的标准分析在中心位置使用所谓的“潜在类别模型”软件(Van Putten等人,《细胞分析》14:86 - 96,1993)进行。该软件提供了一种自动的、“无设门”的淋巴细胞免疫表型分析方法,并采用每次分析运行前定义的固定FL标记设置。在研讨会1中,基于参考仪器获得的结果,为所有仪器在相同的直方图通道中设置这些标记。与使用当地方案进行数据分析相比,统一仪器设置后获取的列表模式数据的标准分析仅使结果的实验室间差异降低了13%。仪器设置的标准方案导致相对较强的FL信号定位统一,但未染色细胞在FL直方图尺度上的定位存在差异。因此,标准FL标记设置对某些仪器不合适。所以,在研讨会2中,使用以MESF单位(等效可溶性荧光染料分子)表示的一系列预定义FL强度的FITC或PE标记微珠对仪器对FITC和PE信号的响应进行校准。该方法允许基于所有仪器以MESF单位表示的相同FL1和FL2强度进行标准分析的FL标记定位。与使用当地方案进行数据分析相比,统一仪器设置并校准FL标记设置后获取的列表模式数据的标准分析使实验室间差异降低了43%。我们得出结论,使用固定FL标记设置的标准列表模式数据分析可降低流式细胞术检测预染色PBMC结果的实验室间差异,前提是仪器已以统一方式设置,并且FL标记已基于对每种仪器对相应FL信号的响应进行校准的基础上进行了标准化。