A comparison of shimming techniques for optimizing fat suppression in MR mammography.

作者信息

Takatsu Yasuo, Nishiyama Kengo, Miyati Tosiaki, Miyano Hideto, Kajihara Mariko, Akasaka Thai

机构信息

Department of Radiology, Osaka Red Cross Hospital, 5-30 Fudegasaki, Tennouji-ku, Osaka 543-8555, Japan.

出版信息

Radiol Phys Technol. 2013 Jul;6(2):486-91. doi: 10.1007/s12194-013-0224-3. Epub 2013 Jun 1.

Abstract

We evaluated the degree of inhomogeneities of fat suppression using the fully automated three-dimensional breast shimming technique (Image Based-Smart: IB-Smart) and manual setting of a rectangular parallelepiped shim (volume shimming) in MR mammography. Information on breast shape was collected from 9 patients whose images were insufficiently fat-suppressed. A breast phantom made of a thermoplastic sheet was used. Shimming of the magnetic field was done with IB-Smart and various dimensions of volume shims: the anterior to posterior/right to left/head to foot directions were set to 75-150/150-350/50-150 mm. The volumes of inhomogeneously suppressed fat were measured. The calculated volume with inhomogeneous fat suppression with use of IB-Smart was 13.3 × 10(4) mm(3). The smallest volume of inhomogeneous fat suppression with volume shimming was 5.4 × 10(4) mm(3) when the anterior-posterior/right-left/head-foot directions were set to 75/350/50 mm. Our results show that using optimized dimensions of volume shims enables better fat suppression than does IB-Smart.