使用中耳假体和耳蜗植入电极阵列植入后,对人体颞骨标本进行数字容积断层扫描和高分辨率计算机断层扫描的放射学检查。

Radiologic examinations in human temporal bone specimens using digital volume tomography and high-resolution computed tomography after implantation of middle ear prosthesis and cochlear implant electrode array.

机构信息

Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Cologne, Germany.

出版信息

Otol Neurotol. 2013 Sep;34(7):1321-8. doi: 10.1097/MAO.0b013e31828bb73c.

Abstract

INTRODUCTION

Adequate imaging of the middle ear and its surrounding structures is an essential preoperative part in ear surgery. In the past, the main disadvantage of computed tomography (CT) scanners was the decreased diagnostic quality due to metallic artifacts. Furthermore, these scanners showed mostly an inferior image quality compared to digital volume tomography (DVT) in the temporal bone. It was the aim of this experimental study to compare the image quality of a state-of-the-art dual-source 2 × 128-slice CT scanner to DVT in temporal bone specimen.

MATERIALS AND METHODS

Metallic prosthesis (PORP, TORP, stapes piston, cochlear implant electrode array) and an autologous incus were implanted in temporal bone specimen to analyze the diagnostic quality concerning the characterization of anatomic structures. Three further temporal bones were scanned without any preparation. Independently, 2 otologists and 2 radiologists scored the image quality of defined anatomic structures, using a range from 4 (excellent) to 0 (no diagnostic value).

RESULTS

The general score for DVT was 2.67, whereas CT reached a score of 2.76. The diagnostic value for hard contrast objects was 3.0 for DVT and 2.9 for CT, whereas the score for soft tissue was 1.1 for DVT and 2.3 for CT. Almost no quality reduction of the display of anatomic structures caused by metallic artifacts could be detected in both diagnostic modalities.

DISCUSSION

Both DVT and high-resolution CT allow good overall image quality in temporal bones, with discrete advantages for digital volume tomographic scans in terms of the image quality of hard contrast objects like bony structures or metallic implants.

摘要

简介

中耳及其周围结构的充分成像,是耳部手术的一个重要术前环节。在过去,计算机断层扫描(CT)的主要缺点是由于金属伪影导致诊断质量下降。此外,与颞骨的数字容积断层扫描(DVT)相比,这些扫描仪的图像质量大多较差。本实验研究的目的是比较最先进的双源 2×128 层 CT 扫描仪与颞骨标本 DVT 的图像质量。

材料与方法

将金属假体(PORP、TORP、镫骨活塞、耳蜗植入电极阵列)和自体砧骨植入颞骨标本中,分析其在解剖结构特征方面的诊断质量。另外三个颞骨标本未经任何准备进行扫描。2 名耳科医生和 2 名放射科医生分别独立对定义的解剖结构的图像质量进行评分,使用 4 分(优秀)至 0 分(无诊断价值)的评分范围。

结果

DVT 的总体评分为 2.67,而 CT 得分为 2.76。DVT 对硬对比物的诊断价值为 3.0,CT 为 2.9,而 DVT 对软组织的诊断价值为 1.1,CT 为 2.3。在这两种诊断方式中,几乎都没有发现由于金属伪影导致的解剖结构显示质量降低。

讨论

DVT 和高分辨率 CT 都能在颞骨中获得良好的整体图像质量,在硬对比物体(如骨结构或金属植入物)的图像质量方面,数字容积断层扫描具有明显优势。