基于手套的记录系统与标准方法的十二导联心电图比较。

Comparison of twelve-lead electrocardiogram using a glove-based recording system with standard methodology.

机构信息

University of Chicago Medical Center, Chicago, IL, USA.

出版信息

Am J Cardiol. 2013 Sep 15;112(6):895-903. doi: 10.1016/j.amjcard.2013.05.018. Epub 2013 Jun 4.

Abstract

We tested the clinically relevant diagnostic accuracy of a new electrocardiographic (ECG) recording system incorporating all 10 ECG electrodes in a single-size glove worn on the patient's left arm and placed on the chest. The PhysioGlove (PG) was designed to allow fast, reproducible, electrode placement with only minimal training. The American College of Cardiology/American Heart Association ECG recording guidelines and others have repeatedly highlighted the unacceptable progressive deterioration in ECG accuracy mainly resulting from a performer's lack of proficiency and diligence, leading to frequent electrode misplacement. We studied 428 consenting adult patients with a broad spectrum of anthropomorphic characteristics and ECG and cardiovascular pathologic entities. The chest girth was measured to ascertain the single-size PG clinical diagnostic accuracy in ≥90% of this patient population. For each patient, a PG and standard-cable electrocardiogram were consecutively recorded and interpreted by experienced electrocardiographers. The study included 3 phases: phase 1, run-in (n = 120); phase 2, comparative diagnostic accuracy (n = 208); and phase 3, randomized, blinded, diagnostic accuracy (n = 100). Of the entire study population (n = 428), 92% fit the chest girth range of 85 to 118 cm (34 to 47 in.), representing the reference standard clinical diagnostic PG chest girth range. The phase 2 PG diagnostic accuracy was 91.3% for entire chest girth range and 95.7% for the 89.4% of patients with a chest girth within the reference range. The mean PG diagnostic accuracy in phase 3 was 93% (95% confidence interval 89% to 95%). In conclusion, compared with standard-cable electrocardiograms, the PG demonstrated excellent diagnostic accuracy (93% to 95.7%) in ≥90% of a typical western adult patient population. The PG's ease of use and minimal training requirements offer a promising tool to markedly improve ECG clinical diagnostic accuracy in most adult western patients.

摘要

我们测试了一种新的心电图(ECG)记录系统的临床相关诊断准确性,该系统将 10 个心电图电极整合在单个尺寸的手套中,戴在患者的左臂上并放置在胸部。PhysioGlove(PG)旨在允许快速、可重复地进行电极放置,只需进行最少的培训。美国心脏病学会/美国心脏协会的 ECG 记录指南和其他指南反复强调,由于表演者缺乏熟练程度和勤奋程度,心电图准确性逐渐恶化,这主要导致电极经常放置不当。我们研究了 428 名同意的成年患者,他们具有广泛的人体特征和心电图及心血管病理实体。测量了胸围,以确定单个尺寸 PG 在≥90%的患者人群中的临床诊断准确性。对于每个患者,连续记录并由经验丰富的心电图医师解释 PG 和标准电缆心电图。该研究包括 3 个阶段:第 1 阶段,试运行(n=120);第 2 阶段,比较诊断准确性(n=208);第 3 阶段,随机、盲法、诊断准确性(n=100)。在整个研究人群(n=428)中,92%的人符合 85 至 118cm(34 至 47 英寸)的胸围范围,代表参考标准临床诊断 PG 胸围范围。整个胸围范围的第 2 阶段 PG 诊断准确性为 91.3%,89.4%的患者胸围在参考范围内的诊断准确性为 95.7%。第 3 阶段的 PG 平均诊断准确性为 93%(95%置信区间 89%至 95%)。总之,与标准电缆心电图相比,PG 在≥90%的典型西方成年患者人群中表现出出色的诊断准确性(93%至 95.7%)。PG 的易用性和最小的培训要求为显著提高大多数西方成年患者的心电图临床诊断准确性提供了一个很有前途的工具。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验