David Vinoo Philip, Stead Richard E, Vernon Stephen A
Department of Ophthalmology, Nottingham University Hospitals NHS Trust, Queens Medical Centre, Nottingham, United Kingdom.
Optom Vis Sci. 2013 Jul;90(7):691-9. doi: 10.1097/OPX.0b013e318297da45.
To assess the repeatability of Goldmann-correlated intraocular pressure (IOPG), corneal-compensated IOP (IOPCC), corneal hysteresis (CH), and the corneal resistance factor (CRF) obtained with the Ocular Response Analyzer (ORA) in normal eyes and to determine whether any differences in corneal biomechanical parameters or their repeatability exist between the sexes.
A prospective observational study assessing 100 normal adults (50 men and 50 women; median age, 54.5 years). Comparison of ORA parameters measured in both eyes in three sets of four consecutive readings by one examiner within a 30-minute period.
The mean values of the ORA parameters assessed, the intraclass correlation coefficient (ICC) and the coefficient of repeatability (CR), were as follows for the right eye (n = 100): IOPG, 16.2 ± 3.3 mm Hg (ICC, 0.96; CR, 6.37); IOPCC, 16.2 ± 3.2 (ICC, 0.94; CR, 6.29); CRF, 10.9 ± 1.9 (ICC, 0.91; CR, 3.62); CH, 10.6 ± 1.7 (ICC, 0.94; CR, 3.37). The ICC between the readings was excellent (>0.9) for all the ORA parameters in both sexes, with the exception of one group (female CRF ICC, 0.86). Coefficient of repeatability of the instrument satisfies the British Standards Institution criteria for repeatability. There were no significant differences between corneal biomechanical factors and IOP results for men and women.
Corneal biomechanical parameters measured by the ORA provide repeatable results in normal eyes, with no significant difference between the sexes.
评估正常眼中使用眼反应分析仪(ORA)获得的戈德曼相关眼压(IOPG)、角膜补偿眼压(IOPCC)、角膜滞后(CH)和角膜阻力因子(CRF)的可重复性,并确定两性之间角膜生物力学参数或其可重复性是否存在差异。
一项前瞻性观察性研究,评估100名正常成年人(50名男性和50名女性;中位年龄54.5岁)。由一名检查者在30分钟内对两眼进行三组连续四次读数,比较ORA参数。
右眼(n = 100)的ORA参数评估均值、组内相关系数(ICC)和可重复性系数(CR)如下:IOPG,16.2±3.3 mmHg(ICC,0.96;CR,6.37);IOPCC,16.2±3.2(ICC,0.94;CR,6.29);CRF,10.9±1.9(ICC,0.91;CR,3.62);CH,10.6±1.7(ICC,0.94;CR,3.37)。除一组(女性CRF的ICC为0.86)外,两性所有ORA参数读数之间的ICC均极佳(>0.9)。仪器的可重复性系数符合英国标准协会的可重复性标准。男性和女性的角膜生物力学因素和眼压结果之间无显著差异。
ORA测量的角膜生物力学参数在正常眼中提供了可重复的结果,两性之间无显著差异。