Suppr超能文献

在放射治疗期间使用连续输注泵。

Use of continuous infusion pumps during radiation treatment.

机构信息

Cancer Care Ontario, Toronto, Canada.

出版信息

J Oncol Pract. 2013 Mar;9(2):107-11. doi: 10.1200/JOP.2012.000717.

Abstract

INTRODUCTION

Despite increasing chemoradiotherapy treatment, there is a paucity of information regarding the effects of radiation exposure on ambulatory infusion pumps used to deliver chemotherapy or other essential medications. The aim of this overview is to present the available evidence on this subject, heighten awareness within the clinical community, provide considerations for minimizing possible negative effects on patient care, and encourage the monitoring of infusion devices after exposure to radiation or electromagnetic interference.

METHODS

Published literature was systematically searched using MEDLINE and EMBASE; gray literature was searched using Google and an environmental scan of relevant Web sites. A multidisciplinary working group reviewed the compiled evidence, and a draft of the document was sent to health professionals from various disciplines for an external review.

RESULTS

Four reports and three manufacturer device alerts were identified that suggest a risk of pump malfunction as a result of radiation exposure. The estimated cumulative dose at which pump failure has been reported ranges from 28.5 to 42 Gy; however, additional clinical investigations should be undertaken. Pump relocation, pump shielding, and assessment of the pump after radiation exposure are most commonly suggested to minimize pump malfunction related to radiation exposure. A list of additional considerations is offered for those developing institution specific policies and procedures based on the available evidence and expert consensus.

CONCLUSION

The varied and unpredictable results of radiation exposure on infusion devices suggest that additional testing should be carried out to determine the limits of dose exposure and to raise awareness around this patient safety issue.

摘要

简介

尽管化学疗法和放射疗法的治疗有所增加,但关于辐射暴露对用于输送化疗或其他必需药物的便携式输注泵的影响的信息仍然很少。本综述的目的是介绍有关该主题的现有证据,提高临床界的认识,为尽量减少对患者护理可能产生的负面影响提供考虑因素,并鼓励在输注设备暴露于辐射或电磁干扰后对其进行监测。

方法

使用 MEDLINE 和 EMBASE 系统地搜索已发表的文献;使用 Google 和相关网站的环境扫描搜索灰色文献。一个多学科工作组审查了汇集的证据,并将文件草案发送给来自不同学科的卫生专业人员进行外部审查。

结果

确定了四项报告和三项制造商设备警报,表明辐射暴露会导致泵故障的风险。据报道,泵故障的估计累积剂量范围为 28.5 至 42 Gy;然而,应该进行更多的临床研究。最常建议重新安置泵、对泵进行屏蔽以及在辐射暴露后评估泵,以尽量减少与辐射暴露相关的泵故障。根据现有证据和专家共识,还提供了其他一些注意事项,以供那些根据具体机构制定政策和程序的人参考。

结论

辐射对输注设备的影响结果各不相同且不可预测,因此应进行更多测试,以确定剂量暴露的极限,并提高对这一患者安全问题的认识。

相似文献

3
Selecting infusion devices for use in ambulatory care.选择用于门诊护理的输液设备。
Am J Health Syst Pharm. 1996 Apr 15;53(8):868-77. doi: 10.1093/ajhp/53.8.868.
7
8
Ambulatory chemotherapy: Past, present, and future.门诊化疗:过去、现在和未来。
J Oncol Pharm Pract. 2021 Jun;27(4):962-973. doi: 10.1177/1078155220985916. Epub 2021 Jan 18.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验