Department of Ophthalmology, Saarland University Medical Center, Homburg, Germany.
Curr Eye Res. 2013 Sep;38(9):910-6. doi: 10.3109/02713683.2013.788722. Epub 2013 Jul 10.
The OA-1000 (Tomey, Japan) is a new optical biometer, which allows measurements of axial length (AL), anterior chamber depth (ACD) and corneal thickness (CT) due to partial coherence interferometry (PCI) technology. The aim of this study was to compare the OA-1000 results with those obtained with the IOLMaster and contact applanation A-scan ultrasonography.
We examined 133 eyes of 75 patients with age related cataract. Mean age was 72.0 ± 9.5 years. Biometry measurements of AL and ACD were performed with the Tomey OA-1000 based on PCI, the IOLMaster based on PCI (AL) and slit projection (ACD), and the Tomey AL-3000 based on contact applanation A-scan ultrasonography.
Mean AL using the IOLMaster was 23.21 ± 1.08 mm, using the AL-3000 was 22.79 ± 1.04 mm, using the OA-1000 it was 22.97 ± 1.1 mm. Mean ACD using the IOLMaster was 2.99 ± 0.41 mm, using the OA-1000 3.40 ± 0.46 mm, using the Tomey AL-3000 it was 2.93 ± 0.43 mm. Mean difference between the AL and ACD measured with the OA-1000 and the IOLMaster was 0.22 ± 0.047 mm and 0.40 ± 0.33 mm, between OA-1000 and the AL-3000 it was 0.19 ± 0.23 mm and 0.47 ± 0.33 mm, and between IOLMaster and AL-3000 it was 0.42 ± 0.23 and 0.09 ± 0.36 mm. For AL the correlation coefficient R between IOLMaster and OA-1000 was 0.999, between IOLMaster and AL-3000 it was 0.976, between AL-3000 and OA-1000 it was 0.977. For ACD R between IOLMaster and OA-1000 was 0.735, between IOLMaster and AL-3000 it was 0.822, between AL-3000 and OA-1000 it was 0.716 (all p < 0.001).
Compared with other used clinical instruments the OA-1000 generates accurate results. Although differences were found, the OA-1000 provided results that correlated well with the values of IOLMaster and AL-3000.
OA-1000(日本托美)是一种新型的光学生物测量仪,采用部分相干干涉测量技术(PCI),可测量眼轴长度(AL)、前房深度(ACD)和角膜厚度(CT)。本研究旨在比较 OA-1000 与 IOLMaster 和接触式压平 A 扫描超声仪的测量结果。
我们对 75 例年龄相关性白内障患者的 133 只眼进行了检查。平均年龄为 72.0±9.5 岁。使用托美基于 PCI 的 Tomey OA-1000、基于 PCI(AL)和裂隙投影(ACD)的 IOLMaster 以及基于接触式压平 A 扫描超声的托美 AL-3000 对 AL 和 ACD 进行生物测量。
IOLMaster 测量的平均 AL 为 23.21±1.08mm,AL-3000 为 22.79±1.04mm,OA-1000 为 22.97±1.1mm。IOLMaster 测量的平均 ACD 为 2.99±0.41mm,OA-1000 为 3.40±0.46mm,托美 AL-3000 为 2.93±0.43mm。OA-1000 和 IOLMaster 测量的 AL 和 ACD 之间的平均差值为 0.22±0.047mm 和 0.40±0.33mm,OA-1000 和 AL-3000 之间的平均差值为 0.19±0.23mm 和 0.47±0.33mm,IOLMaster 和 AL-3000 之间的平均差值为 0.42±0.23mm 和 0.09±0.36mm。对于 AL,IOLMaster 和 OA-1000 之间的相关系数 R 为 0.999,IOLMaster 和 AL-3000 之间的 R 为 0.976,AL-3000 和 OA-1000 之间的 R 为 0.977。对于 ACD,IOLMaster 和 OA-1000 之间的 R 为 0.735,IOLMaster 和 AL-3000 之间的 R 为 0.822,AL-3000 和 OA-1000 之间的 R 为 0.716(均 P<0.001)。
与其他临床仪器相比,OA-1000 产生的结果准确。尽管存在差异,但 OA-1000 提供的结果与 IOLMaster 和 AL-3000 的值相关性良好。