Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes.

机构信息

Department of Ophthalmology, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Xuhui District, Shanghai, China.

出版信息

PLoS One. 2018 Feb 27;13(2):e0193023. doi: 10.1371/journal.pone.0193023. eCollection 2018.

Abstract

PURPOSE

To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster.

METHODS

In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively.

RESULTS

Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively.

CONCLUSION

Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.

摘要

目的

评估 Tomey OA-2000 生物测量仪测量的眼部参数的精密度(重复性和再现性),并将其与 IOLMaster 测量的参数进行比较。

方法

本前瞻性研究纳入了 108 名健康受试者的右眼。两名观察者使用 Tomey OA-2000 连续扫描 3 次,同一次序中,一名观察者使用 IOLMaster(版本 5.4.4.0006)进行测量。大约 1 周后,一名观察者使用 Tomey OA-2000 进行了 3 次扫描。分析 Tomey OA-2000 和 IOLMaster 测量的眼轴(AL)、中央角膜厚度(CCT)、前房深度(ACD)、晶状体厚度(LT)、角膜曲率计读数、瞳孔直径(PD)和角膜直径(CD)值。计算变异系数(CoV)、组内相关系数(ICC)、个体内标准差(Sw)和 2.77Sw 来评估重复性和再现性。分别使用配对 t 检验和 Bland-Altman 图分析两种设备测量的参数的差异和一致性。

结果

Tomey OA-2000 生物测量仪测量的 AL、CCT、ACD、LT、Kf、Ks、Km、PD 和 CD 值的观察者内重复性、观察者间重复性和观察者间可重复性的 CoV 均小于 1%,除 PD 外,ICC 均大于 0.97,除 PD 和 CD 外。Tomey OA-2000 测量的 AL、Kf、Ks、Km 和 CD 值分别低 0.058±0.094mm、0.088±0.150 屈光度(D)、0.163±0.170D、0.127±0.117D 和 0.171±0.217mm(均 P<0.05)。然而,两种设备测量的 ACD 值相当(P=0.169)。AL、ACD、CD 和所有角膜曲率计读数的 95%一致性界限(LoA)均不超过 0.24mm、0.14mm、0.60mm 和 0.5D。

结论

除 PD 和 CD 外,Tomey OA-2000 测量的眼部参数高度可重复和可再现。在健康眼中,Tomey OA-2000 与 IOLMaster 测量的眼部参数除 CD 值外,具有良好的一致性。