Lachapelle P, Blain L
Dept. of Ophthalmology, McGill University, Montreal Children's Hospital, Québec, Canada.
J Neurosci Methods. 1990 Jun;32(3):245-9. doi: 10.1016/0165-0270(90)90147-8.
A Storz model E4110 infant speculum was modified to record electroretinograms (ERGs) from anesthetized rabbits. The electrode combines the advantage of a corneal contact lens electrode (with blepharostat) without its major disadvantage, namely: corneal abrasion. Addition of i.v. tubings glued to both arms of the speculum allows for constant wetting of the eyeball with 0.9% NaCl. The latter not only prevents the eye (and cornea) from drying, but also favors the recording of reproducible ERGs with a noise level comparable to ERGs recorded with a corneal contact lens electrode.