Anal Chem. 2013 Sep 17;85(18):8757-63. doi: 10.1021/ac401857a. Epub 2013 Aug 27.
The stability of the lipid concentration levels in shotgun lipidomics analysis was tracked over a period of 3.5 years. Concentration levels in several lipid classes, such as phospholipids, were determined in human plasma lipid extracts. Impact of the following factors on the analysis was investigated: sample amount, internal standard amount, and sample dilution factor. Moreover, the reproducibility of lipid profiles obtained in both polarity modes was evaluated. Total number of samples analyzed was approximately 6800 and 7300 samples in negative and positive ion modes, respectively, out of which 610 and 639 instrument control samples were used in stability calculations. The assessed shotgun lipidomics approach showed to be remarkably robust and reproducible, requiring no batch corrections. Coefficients of variation (CVs) of lipid mean concentration measured with optimized analytical parameters were typically less than 15%. The high reproducibility indicated that no lipid degradation occurred during the monitored time period.
在长达 3.5 年的时间里,研究人员跟踪了 shotgun 脂质组学分析中脂质浓度水平的稳定性。在人血浆脂质提取物中,研究人员测定了几种脂质类别的浓度水平,如磷脂。研究考察了以下因素对分析的影响:样本量、内标量和样本稀释因子。此外,还评估了在两种极性模式下获得的脂质图谱的重现性。在负离子模式和正离子模式下,分别分析了约 6800 和 7300 个样本,其中 610 和 639 个仪器控制样本用于稳定性计算。评估的 shotgun 脂质组学方法表现出显著的稳健性和重现性,无需批次校正。使用优化的分析参数测量的脂质平均浓度的变异系数(CV)通常小于 15%。高重现性表明在监测期间没有发生脂质降解。