[Validation of the Holter method in detecting myocardial ischemia using an exercise test simultaneously].

作者信息

Rayo I, Marín Huerta E, Muela A, Moro C, Soler M

机构信息

Servicio de Cardiología, Hospital Ramón y Cajal, Madrid.

出版信息

Rev Esp Cardiol. 1990 May;43(5):316-22.

PMID:2392611
Abstract

UNLABELLED

To evaluate the reliability of Holter monitoring in reproducing myocardial ischemic changes, 110 patients (90 males, 20 females; age range: 14-74 years) underwent a Bruce protocol treadmill exercise test. An electrocardiogram was recorded simultaneously with a two-channel modulated Holter recorder (frequency response: 0.05-100 Hz) with bipolar CM-V3 and CM-V5 leads and by a conventional 12-lead system. An ischemic ST-segment change was defined as 1 mm or more ST-segment depression lasting more than 0.08 sec after the J point. Results were concordant in 101 patients, 36 with both positive and 65 with both negative responses. Eight false negative and one false positive Holter ischemic episodes occurred. This yielded an accuracy (expressed as sensitivity, specificity, positive and negative predictive values) of 81.8%, 98.5%, 97.3% and 89.0%, respectively. There was a good correlation between the maximal ST-segment depression (r = 0.57; p less than 0.001), duration of ischemia (r = 0.89; p less than 0.001), heart rate at the onset of the ischemic episode (r = 0.91; p less than 0.001) and maximal heart rate (r = 0.98; p less than 0.001).

CONCLUSIONS

Two-channel amplitude modulated Holter recording system with bipolar CM-V3 and CM-V5 leads can reliably reproduce ST-segment changes.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验