Fetal blood flow velocimetry by phase-contrast MRI using a new triggering method and comparison with Doppler ultrasound in a sheep model: a pilot study.

作者信息

Schoennagel Bjoern P, Remus Chressen C, Yamamura Jin, Kording Fabian, Tavares de Sousa Manuela, Hecher Kurt, Fischer Roland, Ueberle Friedrich, Boehme Matthias, Adam Gerhard, Kooijman Hendrik, Wedegaertner Ulrike

机构信息

Department of Diagnostic and Interventional Radiology, University Medical Center Hamburg-Eppendorf, Martinistraße 52, 20246, Hamburg, Germany,

出版信息

MAGMA. 2014 Jun;27(3):237-44. doi: 10.1007/s10334-013-0397-0. Epub 2013 Aug 10.

Abstract

OBJECT

We present the first study demonstrating the feasibility of antenatal blood flow velocimetry performing ECG triggered phase-contrast (PC)-MRI in the fetal aorta by using a newly developed Doppler ultrasound trigger.

MATERIALS AND METHODS

Five pregnant sheep carrying singleton fetuses (gestational age 121 days) were anesthetized to undergo fetal 2D PC-MRI in the fetal descending aorta (1.5 T) using a newly developed MR-compatible Doppler ultrasound trigger for fetal cardiac triggering. Inter-operator variability was assessed for PC-MR measurements and reproducibility was tested by repeated scans in one fetus. Inter-modality comparison was performed by Doppler ultrasound velocimetry.

RESULTS

Fetal cardiac triggering was possible in all examinations. PC-MR velocimetry revealed a mean inter-operator variability of 3 ± 5%. Average peak systolic flow velocities of 62.5 ± 4.4 cm/s were in good agreement with Doppler ultrasound measurements of 62.0 ± 9.2 cm/s (p (Lord's U test) ≫ 0.05).

CONCLUSION

Fetal PC-MR velocimetry was successfully performed using the newly developed MR-compatible Doppler ultrasound trigger for intrauterine fetal cardiac triggering, demonstrating high inter-operator and inter-modality agreement. This new method has the high potential for alternative assessment of hemodynamic decompensation of the fetal circulation.

摘要

目的

我们开展了首项研究,通过使用新开发的多普勒超声触发器,展示在胎儿主动脉中进行心电图触发相位对比(PC)-MRI的产前血流测速的可行性。

材料与方法

对5只怀有单胎胎儿(妊娠121天)的怀孕绵羊进行麻醉,使用新开发的用于胎儿心脏触发的磁共振兼容多普勒超声触发器,在胎儿降主动脉(1.5T)中对胎儿进行二维PC-MRI检查。评估了PC-MR测量的操作者间变异性,并通过对一只胎儿进行重复扫描测试了可重复性。通过多普勒超声测速法进行了模态间比较。

结果

所有检查中均可实现胎儿心脏触发。PC-MR测速显示操作者间平均变异性为3±5%。平均收缩期峰值流速为62.5±4.4cm/s,与多普勒超声测量值62.0±9.2cm/s高度一致(p(洛德U检验)≫0.05)。

结论

使用新开发的用于子宫内胎儿心脏触发的磁共振兼容多普勒超声触发器成功进行了胎儿PC-MR测速,显示出较高的操作者间和模态间一致性。这种新方法在替代评估胎儿循环血流动力学失代偿方面具有很大潜力。