Accelerated phase-contrast MR imaging: comparison of k-t BLAST with SENSE and Doppler ultrasound for velocity and flow measurements in the aorta.

作者信息

Stadlbauer Andreas, van der Riet Wilma, Globits Sebastian, Crelier Gerard, Salomonowitz Erich

机构信息

MR Physics Group, Department of Radiology, Landesklinikum St. Poelten, Austria.

出版信息

J Magn Reson Imaging. 2009 Apr;29(4):817-24. doi: 10.1002/jmri.21706.

Abstract

PURPOSE

To evaluate differences in velocity and flow measurements in the aorta between accelerated phase-contrast (PC) magnetic resonance imaging (MRI) using SENSE and k-t BLAST and in peak velocity to Doppler ultrasound.

MATERIALS AND METHODS

Two-dimensional PC-MRI perpendicular to the ascending and descending aorta was performed in 11 volunteers using SENSE (R = 2) and k-t BLAST (2-, 4-, 6-, and 8-fold). Peak velocity, mean velocity, and stroke volume of the accelerated PC-MRI experiments were correlated. Peak velocities were compared to Doppler ultrasound.

RESULTS

All acceleration techniques showed significant correlations for peak velocity with Doppler ultrasound. However, k-t BLAST 6 and 8 showed a significant underestimation. Strong correlations between SENSE and k-t BLAST were found for all three parameters. Significant differences in peak velocity were found between SENSE and all k-t BLAST experiments, but not for 2-fold k-t BLAST in the ascending aorta, and 2- and 4-fold k-t BLAST in the descending aorta. For mean velocity no significant differences were found. Stroke volume showed significant differences for all k-t BLAST experiments in the ascending and for 6- and 8-fold k-t BLAST in the descending aorta.

CONCLUSION

Peak velocity of accelerated PC-MRI correlated with CW Doppler measurements, but high k-t BLAST acceleration factors lead to a significant underestimation. SENSE with R = 2 and 2-fold k-t BLAST are most highly correlated in phase-contrast flow measurements.

摘要

目的

评估使用敏感度编码(SENSE)和k-t BLAST的加速相位对比(PC)磁共振成像(MRI)与多普勒超声在主动脉速度和血流测量方面的差异。

材料与方法

对11名志愿者进行垂直于升主动脉和降主动脉的二维PC-MRI检查,分别使用SENSE(加速因子R = 2)和k-t BLAST(加速因子为2倍、4倍、6倍和8倍)。对加速PC-MRI实验中的峰值速度、平均速度和搏出量进行相关性分析。将峰值速度与多普勒超声测量结果进行比较。

结果

所有加速技术的峰值速度与多普勒超声均显示出显著相关性。然而,k-t BLAST 6倍和8倍加速时出现显著低估。SENSE和k-t BLAST在所有三个参数上均显示出强相关性。SENSE与所有k-t BLAST实验在峰值速度上存在显著差异,但在升主动脉中2倍k-t BLAST以及降主动脉中2倍和4倍k-t BLAST时除外。平均速度方面未发现显著差异。在升主动脉中,所有k-t BLAST实验的搏出量均存在显著差异,在降主动脉中,6倍和8倍k-t BLAST时的搏出量存在显著差异。

结论

加速PC-MRI的峰值速度与连续波多普勒测量结果相关,但高k-t BLAST加速因子会导致显著低估。在相位对比血流测量中,加速因子R = 2的SENSE和2倍k-t BLAST相关性最高。