Stadlbauer Andreas, van der Riet Wilma, Globits Sebastian, Crelier Gerard, Salomonowitz Erich
MR Physics Group, Department of Radiology, Landesklinikum St. Poelten, Austria.
J Magn Reson Imaging. 2009 Apr;29(4):817-24. doi: 10.1002/jmri.21706.
To evaluate differences in velocity and flow measurements in the aorta between accelerated phase-contrast (PC) magnetic resonance imaging (MRI) using SENSE and k-t BLAST and in peak velocity to Doppler ultrasound.
Two-dimensional PC-MRI perpendicular to the ascending and descending aorta was performed in 11 volunteers using SENSE (R = 2) and k-t BLAST (2-, 4-, 6-, and 8-fold). Peak velocity, mean velocity, and stroke volume of the accelerated PC-MRI experiments were correlated. Peak velocities were compared to Doppler ultrasound.
All acceleration techniques showed significant correlations for peak velocity with Doppler ultrasound. However, k-t BLAST 6 and 8 showed a significant underestimation. Strong correlations between SENSE and k-t BLAST were found for all three parameters. Significant differences in peak velocity were found between SENSE and all k-t BLAST experiments, but not for 2-fold k-t BLAST in the ascending aorta, and 2- and 4-fold k-t BLAST in the descending aorta. For mean velocity no significant differences were found. Stroke volume showed significant differences for all k-t BLAST experiments in the ascending and for 6- and 8-fold k-t BLAST in the descending aorta.
Peak velocity of accelerated PC-MRI correlated with CW Doppler measurements, but high k-t BLAST acceleration factors lead to a significant underestimation. SENSE with R = 2 and 2-fold k-t BLAST are most highly correlated in phase-contrast flow measurements.
评估使用敏感度编码(SENSE)和k-t BLAST的加速相位对比(PC)磁共振成像(MRI)与多普勒超声在主动脉速度和血流测量方面的差异。
对11名志愿者进行垂直于升主动脉和降主动脉的二维PC-MRI检查,分别使用SENSE(加速因子R = 2)和k-t BLAST(加速因子为2倍、4倍、6倍和8倍)。对加速PC-MRI实验中的峰值速度、平均速度和搏出量进行相关性分析。将峰值速度与多普勒超声测量结果进行比较。
所有加速技术的峰值速度与多普勒超声均显示出显著相关性。然而,k-t BLAST 6倍和8倍加速时出现显著低估。SENSE和k-t BLAST在所有三个参数上均显示出强相关性。SENSE与所有k-t BLAST实验在峰值速度上存在显著差异,但在升主动脉中2倍k-t BLAST以及降主动脉中2倍和4倍k-t BLAST时除外。平均速度方面未发现显著差异。在升主动脉中,所有k-t BLAST实验的搏出量均存在显著差异,在降主动脉中,6倍和8倍k-t BLAST时的搏出量存在显著差异。
加速PC-MRI的峰值速度与连续波多普勒测量结果相关,但高k-t BLAST加速因子会导致显著低估。在相位对比血流测量中,加速因子R = 2的SENSE和2倍k-t BLAST相关性最高。